04/03
347
LB 444
47
Message
Cause, Notes
Measurement halted
<HALT>
Measurement stopped via digital input
or
Measurement stopped because count rate has
exceeded or not reached count rate threshold.
Rate Overflow
Count rate is > 520,000 (e.g. empty measuring
path, detector error)
Power Fail
Complete failure or drop of power supply below
tolerance level.
No Detector Rate *)
Detector supplies no pulses for at least 60 s.
Wrong HV *)
HV reference voltage of detector too high or
too low.
Measurement is stopped when this error occurs
and has to be restarted by pushing the <
run
>
button.
Measurement is not stopped. Relay “Failure” is
de-energized.
Error temperature measurement
Temperature input faulty (see “Temperature
Compensation” and “Service Instructions”).
Power failure > 1 month *)
Enter New Date
After a long period of power failure you have to
check and, if necessary, correct the date to en-
sure correct decay compensation.
Even if the date is correct, you have to enter
one digit of the year and confirm the input with
<
enter
>.
*) When these error messages are displayed, the measurement stops automatically and has to be restarted by
pushing the <
run
> button
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...