8.
CALIBRATION
Calibration is carried out as follows:
The counts supplied by the detector are read into memory at various product densities or
concentrations. At the same time, the product temperature is automatically stored. The
density or concentration values (analysis values) determined in the lab are entered in
another memory. The diagram below illustrates the correlation of memories.
Count rates:
Rate:
Density/
Concentration
(Lab values):
Product
temperature:
1: 3012
1.234
25
2: 2801
1.322
22
3: 2567
1.381
23
4: 2711
1.363
...
...
...
...
...
10: 0
0
0
Calibration points should be distributed fairly evenly over the entire measuring
range.
Some calibration points may lie outside the measuring range (e.g. water). How-
ever, calibration with
all
points lying outside the measuring range is
not advisable
.
8.1.1
Calibration Modes
The calibration mode has to be entered in the
Calibrate
menu, submenu
Calculate
. Five
calibration modes are available.
Calibrate mode 1
: One-point calibration
Select
one
and enter the
absorption coefficient
Select this calibration mode if
•
only one value pair is available
or
•
only provisional calibration is to be performed.
Calibrate mode 2
: Linear curve fit
Select
lin.
Chose this calibration if only two value pairs are available.
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...