04/03
347
LB 444
57
8.1.2
One-Point Calibration
For one-point calibration one needs a value pair (count rate and density or concentration
value) and the absorption coefficient. The absorption coefficient for common products is
listed in the enclosed table (see Appendix). For products not listed there, enter the fol-
lowing values for the unit g/cm
3
.
Cs-137:
Frontal (axial)
irradiation
Lateral
irradiation
Crystal size
40/35
25/25
50/50 All
crystal
sizes
Absorption coefficient µ
- 0.0664
- 0.06
-0.057
Co-60:
Frontal irradiation: µ = - 0.0478
Lateral irradiation: µ = - 0.0403
For other units the absorption coefficients have to be converted as follows:
µ
µ
µ
ρ
ρ
ρ
ρ
x
C
C
C
=
=
−
−
*
*
max
min
max
min
∆ρ
∆
Example 1:
Isotope: Cs-137
Crystal: 40/35; frontal irradiation
ρ
max
= 1.098
g/cm
3
: C
ρ
max = 20%
ρ
min
= 1.045
g/cm
3
: C
ρ
min = 10%
µ
%
.
*
.
.
.
= −
−
−
= −
0 0664
1098 1045
20 10
0 00352
C
ρ
= Concentration in % at respective density value
Example 2:
ρ
max
= 0,95 g/cm
3
: C
ρ
max
= 30 %
ρ
min
= 0,91 g/cm
3
: C
ρ
min
= 50 %
000133
.
0
50
30
91
.
0
95
.
0
*
0664
.
0
%
=
−
−
−
=
µ
Input: + 1.3300e - 04
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...