8.5.2
Multiplication Factor
Each measured value is multiplied by the value stored as
Factor
. This allows you to chan-
ge the gradient of the calibration curve.
Example:
Measuring range 1.1 - 1.3 g/cm
3
If you enter 1.1 at
Factor
, 1.21 g/cm
3
will be displayed instead of 1.1 g/cm
3
. Instead of
1.3 g/cm
3
the value 1.43 g/cm
3
is now displayed. Thus even the low point of the measur-
ing range is increased, provided it is not at 0.
Remedy:
This can be corrected by entering the appropriate data under
Offset
and
Factor
.
Example:
For 1.1 g/cm
3
the indicated value is correct; instead of 1.2 g/cm
3
the reading should
show 1.25 g/cm
3
.
Input value for
Factor
:
3
3
3
3
actual
actual
nominal
nominal
g/cm
1.1
-
g/cm
1.2
g/cm
1.1
-
g/cm
1.25
=
L
-
H
L
-
H
=
F
= 1.5
Input value for
Offset
:
K
=
L
nominal
- L
actual
F = 1.1 g/cm
3
- 1.1 g/cm
3
· 1.5 g/cm
3
= -0.55
H
= upper value of the measuring range in g/cm
3
L
= lower value of the measuring range in g/cm
3
The same is true if upper and lower point of the measuring range are to be changed by
different values.
Example:
L
actual
= 1.12 g/cm
3
L
nominal
= 1.15 g/cm
3
H
actual
= 1.25 g/cm
3
H
nominal
= 1.3 g/cm
3
F
g cm
g cm
g cm
g cm
=
−
−
13
115
125
112
3
3
3
3
.
/
.
/
.
/
.
/
K = 1.15 g/cm
3
- 1.12 g/cm
3
· 1.1538 = - 0.1423
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...