Service Instructions
10.1
General Safety Precautions
Any time you are working on electrical components, you have to observe the relevant
safety regulations. Please refer to the Safety Summary at the beginning of this operating
manual.
Any work in the direct vicinity of the shielding containing the radioactive source may be
performed only following proper training and/or professional guidance.
10.2
Evaluation Unit LB 444
The evaluation unit is provided with an error control which also monitors the detector
functions. An error code appears on the display if an error is detected. For information
about the error and possible causes please refer to the error code list. In case of hard-
ware errors the evaluation unit has to be replaced.
If no error code is displayed, the electronics is working properly and all measured values
and parameters lie in the normal possible range. Possible function problems must then
have another cause.
“Total Reset” should be performed only in exceptional cases and after consultation with
Berthold.
To perform a “Total Reset”, turn the instrument off; then push the
Clear
button while
you turn the instrument on.
All entered date will be deleted and the default values restored.
Error code list
No. Problem
Cause
1
Rate Overflow
Count rate > 520000cps
2
No Detector Rate
Error in detector or during transmission
3
Wrong HV
HV < 500V or > 1500V, outside control range
6 Detector
temperature
> 65°C
Detector temperature has exceeded > 65°C
7
Trouble with detector
communication
Data transfer between detector and evaluation unit faulty
8 Measurement
stopped
Measurement
stopped by closed contact 22 a/c
9 Detector
temperature
> 80°C
Detector temperature has exceeded > 80°C
10 Missing calibration
Calibration has to be terminated with <
enter
> and
<
more
>
11 Power fail > 1 month
Instrument has been turned off for more than one
month. Enter new date.
12 No input allowed
Device locked by password or wrong password entered
33 Wrong password
Wrong password entered
36 Not enough points
Not enough calibration points available
37 Not steady
Calibration not steady
39 Measuring path not
defined
Measuring path not yet entered
46 Radiating interference
New start with <
run
> or via input 3 (22a/22c)
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...