04/03
347
LB 444
75
49 Liquid density / Solid den-
sity not entered
(Rho.S/RhoL)
Liquid density and/or Solid density not entered
50 I.Max/I.Min value
Current output entered incorrectly
51 Current input error
Current input range entered incorrectly
52 Current output limit value Current output range entered incorrectly
53 Current input error
Current input too high > 20mA
54 Device
not
calibrated
Calibration has to be terminated with <
enter
>
56 Range overflow
Input value outside permissible range
57 Wrong input value
Input value outside permissible range
58 Calibration faulty
Wrong calibration
60 Digital inputs for product
selection allowed
Window for product selection in menu
Parameter
set to
external
.
Problem Possible
Cause
Potential
Solution
Check mains cables
No display
No power supply
Check fuses
Display unreadable Processor error
Note error code
Perform total reset
Replace evaluation unit
Power supply for
detector interrupted
Check cable to detector
No pulse rate
(Error code 2)
Detector is faulty
Replace detector
Shielding not open or
not properly opened
Check lock and secure it in OPEN position
Pulse rate too low
Alignment of useful
beam toward detector
not correct
Correct and optimize alignment
Container
installations
in radiation path
Offset irradiation level
Wall deposits in
container
Remove wall deposits
No or incorrect
result reading
Entry of calibration
values incorrect
Check correlation of count rate and lab values
Output current
incorrect
Entry of calibration
values incorrect
Check correlation of count rate and lab values
Time constant too
small
Increase time constant in the Parameter
menu
Rapid switchover with
Sigma value too small
Disable rapid switchover
Result reading
fluctuates strongly
Pulse rate too low
Check age of source and irradiation level.
Replace detector
Detector stabilization
faulty
Replace detector
Photomultiplier faulty Replace photomultiplier
Result reading
shows drifts
Incorrect isotope
entered
While measurement is running:
enter calibration date
correct isotope
enter current date
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...