04/03
347
LB 444
81
Request Plateau
The value pairs of the plateau measurement may be displayed here and may be plotted
to assess the curve.
Adjust current output
Current 1.8 or 18 mA:
Adjustment of output current to the values 1.8 mA and 18 mA by incrementing or dec-
rementing the numerical value on the display.
Adjust current input
Target value:/Actual value
Adjustment of input current. Feed current of 1.8 mA.
Confirm display with <
enter
>.
RUN appears on the display, the adjustment is running. Push <
more
> to stop the ad-
justment and continue with the next item.
Then feed 18 mA current and confirm with <
enter
> and push <
more
> to continue.