6
2.3 Setting the logger’s properties to start a new measurement.
1. Connect the data logger to the computer.
1
2. Running BSTSoftware on the PC.
3. From the toolbar select Connect.
2
4. Read the logger’s properties.
5. Set the properties.
6. Sync the properties & time. (The PC time will be synchronized to the logger as well.)
7. From the toolbar select Disconnect
8. Unplug the logger from the Computer, and then the logger is in Standby mode
3
.
9. Press OK button on the logger, turn on the Data Logger and start recording.
1
The windows operating system cannot handle USB devices being unplugged and plugged back too
fast. When unplugging the logger, wait for about 5 seconds before plugging it in again. If you unplug
and plug back a device too quickly, the computer may stop recognizing any USB devices on that port.
If this happens you will have to restart the computer. This is a windows USB problem and is not
related to BSTSoftware.
2
If more than one logger is connected to your computer at the same time, the program will ask you to choose
COM port manually.
3
The logger has three modes: 1.LOG: In the mode, the logger samples and records data timely. 2.
Standby: In the mode, the logger stops to sample and record, and the LCD display is OFF. 3. OFF: In
the mode, the logger stops to sample and record, and the LCD display is off. BSTSoftware cannot
connect to the logger also.