should be inserted into the Gen Out output terminal, and the probe grounding clamp
should be clamped on the metal outer ring of the Gen Out output terminal.
Recommended input ~2V@1KHz peak-peak square wave.
3) Press the
[Auto]
button and you should see within a few seconds a square wave
of about 2V peak-to-peak at 1KHz in the display. Repeat the steps to observe CH2.
Probe Check
Safety
When using the probe, keep your fingers behind the guard on the probe body to avoid
electric shock. Do not touch metallic portions of the probe head while it is connected to a
voltage source. Connect the probe to the oscilloscope and connect the ground terminal to
ground before you start any measurements.
Manual Probe Compensation
Upon the first connection of a probe and an input channel, you should manually perform
this adjustment to match the probe to the input channel. Uncompensated or
miscompensated probes may lead to errors or faults in measurement. To adjust the
probe compensation, follow the steps below.
1. Press
Channel
button to enter channel setting menu, Set the Probe option
attenuation in the channel menu to 10X. Set the switch on the probe to 10X and
connect the probe to Channel 1 on the oscilloscope. If you use the probe hook-tip,
the hook end should be removed, the probe pin should be inserted into the Gen Out
output terminal, and the probe grounding clamp should be clamped on the metal
outer ring of the Gen Out output terminal. The oscilloscope with signal generator
function needs to set output signal as 2V @ 1KHz square wave; The Gen Out
terminal of the oscilloscope without signal generator function automatically outputs
2V @ 1KHz square wave. Press the
[Auto]
button.
2. Check the shape of the displayed waveform.
Compensated correctly
Overcompensated
Undercompensated
55
Summary of Contents for 1760DGT/2
Page 1: ...1760DGT 2 Multimetro digitale con oscilloscopio Manuale d uso 1 ...
Page 44: ...1760DGT 2 Digital multimeter with oscilloscope User Manual 44 ...
Page 75: ...8 Connect Gen Out connector to oscilloscope for observation 75 ...
Page 85: ...1760DGT 2 Multimètre numérique avec oscilloscope Manuel de l Utilisateur 85 ...
Page 117: ...8 Connectez le connecteur Gen Out à l oscilloscope pour l observation 117 ...
Page 127: ...1760DGT 2 Digitalmultimeter mit Oszilloskop Benutzerhandbuch 127 ...
Page 159: ...8 Schließen Sie den Gen Out Stecker zur Beobachtung an das Oszilloskop an 159 ...
Page 169: ...1760DGT 2 Multímetro digital con osciloscopio Manual de usuario 169 ...
Page 200: ...8 Conecte el conector Gen Out al osciloscopio para observación 200 ...
Page 210: ...1760DGT 2 Multímetro digital com osciloscópio Manual do usuário 210 ...
Page 241: ...8 Conecte o conector Gen Out ao osciloscópio para observação 241 ...
Page 251: ...1760DGT 2 Digitale multimeter met oscilloscoop Handleiding 251 ...
Page 282: ...8 Sluit de Gen Out connector aan op de oscilloscoop voor observatie 282 ...
Page 292: ...1760DGT 2 Multimetr cyfrowy z oscyloskopem Instrukcja obsługi 292 ...
Page 323: ...8 Podłączyć złącze Gen Out do oscyloskopu w celu obserwacji 323 ...
Page 333: ...1760DGT 2 Digitális multiméter oszcilloszkóppal Használati utasítás 333 ...
Page 365: ...8 Csatlakoztassa a Gen Out csatlakozót az oszcilloszkóppal megfigyelés céljából 365 ...