Better Water LLC; rev. Jul 2019
Page 35 of 80
PB2
– Mediport Portable RO Operator Manual
SET-UP FOR A DIALYSIS TREATMENT
1.
Connect
Inlet Water Hose
(clear-braided hose)
to water source.
2.
Place
Drain Hose
(red tube)
to drain.
- If equipped with a Divert-to-Drain, then also place the Divert-to-
Drain’s Drain Tube to drain.
3.
Connect
Product Hose
(blue tube)
to dialysis machine.
- If equipped with a Divert-to-Drain, the Product Tube will come from it, otherwise it will come
directly from the PB2.
- NOTE: It is important to note that when the PB2 product line is connected to the dialysis
machine and the PB2 is running, the dialysis machine must be pulling water in order for an
accurate reading of the water quality monitor. It will take approximately 15-20 minutes before
optimum water quality is reached. There will be no product pressure or when the product hose is
not connected to the dialysis machine.
4.
Turn
Water Source
to the PB2
ON.
5.
Plug PB2 into an electrical receptacle.
6.
Turn
OPERATE Switch
on the PB2 to the
OPERATE
position.
- If the Low Pressure Alarm sounds press the
LOW PRESSURE ALARM RESET Button
on the
PB2.
- If the PB2 continues to go into Low Pressure Alarm, see
“
Low Pressure Alarm” Section.
7.
Perform chlorine/chloramine check.
- It is highly recommended that before each dialysis treatment a check should be made for
chlorine/chloramine breakthrough which should not exceed 0.1 mg/L.
- If external carbon blocks or tanks are used, the internal carbon block should have been
removed and replaced with a false cartridge.
-
Open the PB2’s cabinet door and obtain a sample of water from the Labcock located on the inlet
solenoid valve.
* The water obtained from this location is pre-treated water only not AAMI standard water.
Do not use this water for bicarb make-up or any other application where AAMI standard
quality water is required.
- Do not proceed unless water results are within the desire range.
8.
Operational Notes:
- Filter-In and Filter-Out pressures should not exceed 15 psi differential. If it does, change the 5
micron filter
and/or the 10” carbon block inside the PB2.
- Product pressure will vary as the dialysis machine uses water. Not to exceed 70 psi.
- Verify readings on the Water Quality Monitor to ensure good water quality. If water quality is
90% rejection or less the PB2 will shut-off and alarm, unless the set-point has been reset.
9.
Once the dialysis treatment is complete the PB2 should be flushed for 15 minutes.
- See
“Flushing the PB2” section
for more information.
- Turn the
OPERATE Switch
to the
ON
position to start the PB2.
- If the PB2 goes into Low Pressure Alarm, press the
LOW PRESSURE ALARM RESET Button
.
If it continues to go into Low Pressure Alarm, see
“
Low Pressure Alarm” Section.
- Turn the
Flush Valve
located on the left side of the PB2 to the
ON
position.
- After 15 minutes of flushing, turn the
Flush Valve
to the
OFF
position.
10.
Turn the
OPERATE Switch
to the
OFF
position.
11.
Turn
Water Source
to the PB2
OFF
, and disconnect hoses.
Summary of Contents for EQROPB01600
Page 1: ...REA 2380 PB2 Mediport Portable RO Operator Manual Better Water LLC rev Jul 2019 ...
Page 20: ...Better Water LLC rev Jul 2019 Page 16 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 36: ...Better Water LLC rev Jul 2019 Page 32 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 56: ...Better Water LLC rev Jul 2019 Page 52 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 80: ...Better Water LLC rev Jul 2019 Page 76 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 82: ...Better Water LLC rev Jul 2019 Page 78 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 84: ...Better Water LLC rev Jul 2019 Page 80 of 80 PB2 Mediport Portable RO Operator Manual ...