Better Water LLC; rev. Jul 2019
Page 61 of 80
PB2
– Mediport Portable RO Operator Manual
TECHNICAL SERVICE BULLETIN
PB1/PB2 Low PSI / Monitor Modification
TSB#
TSB2012001
Date
01/11/12
Page 6 of 6
Verifying Modifications:
After the modification is complete, BEFORE the cabinet is closed and put into service, you must
ensure the accuracy of the work performed for safety and proper operation by doing the following:
1) When in operate mode, the Water Quality Monitor should power ON and remain ON after
pressing the Reset Button.
2) Simulate a low water pressure condition by turning the tap water supply OFF to the RO unit. The
following should happen:
a) Inlet Solenoid Valve should close,
b) pump will shut down,
c) Water Quality Monitor will remain on.
3) Verify the Inlet Water Solenoid Valve is closed by removing red drain line from drain
– after a few
seconds, there should be no constant water flow.
4) After verification of the modification:
a) turn the tap water supply ON,
b) close / replace the control box cover.
IMPORTANT OPERATIONAL NOTE:
After this modification and BEFORE pressing the START button, during
initial startup or filter change, air will need to be purged from the unit
at the labcock on the inside of the cabinet.
Summary of Contents for EQROPB01600
Page 1: ...REA 2380 PB2 Mediport Portable RO Operator Manual Better Water LLC rev Jul 2019 ...
Page 20: ...Better Water LLC rev Jul 2019 Page 16 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 36: ...Better Water LLC rev Jul 2019 Page 32 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 56: ...Better Water LLC rev Jul 2019 Page 52 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 80: ...Better Water LLC rev Jul 2019 Page 76 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 82: ...Better Water LLC rev Jul 2019 Page 78 of 80 PB2 Mediport Portable RO Operator Manual ...
Page 84: ...Better Water LLC rev Jul 2019 Page 80 of 80 PB2 Mediport Portable RO Operator Manual ...