31
Interval between
2 measurements
At least 5 seconds
Clinical repeat precision
Ear: Children, 1 – 5 years: ±0.14 °F (±0.08 °C)
Adults: ±0.13 °F (±0.07 °C)
Forehead: Children, 1 – 5 years: ±0.13 °F (±0.07 °C)
Adults: ±0.14 °F (±0.08 °C)
Measurement units
°Celsius (°C) or °Fahrenheit (°F)
Ambient operating conditions 59 °F – 95 °F (15°C – 35°C) with a relative humidity of up to 85% (non-condensing)
Storage conditions
-13 °F – 131 °F (-25°C – 55°C) with a relative humidity of up to 85% (non-condensing)
Dimensions
38.2 x 138 x 46.5 mm
Weight
90 g including battery
Battery
2 x 1.5V AAA (LR03)
Memory
For 10 measurements
The serial number is located on the device or in the battery compartment.
12. Trouble-shooting
Display
Cause
Solution
The recorded temperature is higher than
1) Ear/Forehead thermometer mode: 109.4 °F (43°C)
2) Object temperature mode: 212 °F (100°C)
Operate the thermometer only between the specified
temperature ranges.
If necessary, clean the sensor tip. In the event of a
repeated error message, contact your retailer or Cus-
tomer Services.
The recorded temperature is lower than
1) Ear/Forehead thermometer mode: 93.2 °F (34°C)
2) Object temperature mode: 32 °F (0°C)
The operating temperature is not in the range from 59
to 95 °F (15°C to 35°C).
Operate the thermometer only between the specified
temperature ranges.