30
12. Trouble-shooting
Display
Cause
Voice output
Solution
The recorded temperature is higher than
1) Ear/Forehead thermometer mode:
109.4 °F (43 °C),
2) Object temperature mode:
212 °F (100 °C).
Measurement error,
please try again!
Operate the thermometer only
between the specifi ed temperature
ranges.
If necessary, clean the sensor tip.
In the event of a repeated error
message, contact your retailer or
Customer Services.
The recorded temperature is lower than
1) Ear/Forehead thermometer mode:
93.2 °F (34 °C),
2) Object temperature mode: 32 °F (0 °C).
Measurement error,
please try again!
The operating temperature is not in the range
from 60.8 °F to 95 °F (16 °C to 35 °C).
Measurement error,
please try again!
Operate the thermometer only
between the specifi ed temperature
ranges.
Summary of Contents for FT 70
Page 9: ......
Page 10: ......
Page 15: ......
Page 24: ......
Page 25: ......
Page 29: ......
Page 38: ......
Page 39: ......
Page 44: ......
Page 52: ......
Page 53: ......
Page 57: ......
Page 66: ......
Page 67: ......
Page 71: ......
Page 80: ......
Page 81: ......
Page 85: ......
Page 94: ......
Page 95: ......
Page 100: ......
Page 110: ......
Page 111: ......
Page 115: ......