26
Beurer GL43 mg/dL
• If the device is used in a dry environment, in particular near synthetic materials (clothes containing
synthetic fibres and carpets, for example), the damaging static charges which occur may cause er-
roneous results.
• Do not use the device near sources of strong electro-magnetic radiation, as this may affect normal
operation.
• It is a good idea to carry out an assessment of the electro-magnetic environment before using the
device commercially.
9. w
hat
if
there
are
problems
?
Display messages on batteries and blood glucose measurement
no.
Cause
Solution
LP
Empty battery.
Replace battery.
Ht
Temperature of the measuring environ-
ment, blood glucose monitor or test
strip above the permitted range.
Repeat the test using a new test strip as soon as
the measuring environment, measuring device
and test strips have reached room temperature
(+20 °C to +26 °C).
Lt
Temperature of the measuring environ-
ment, blood glucose monitor or test
strip below the permitted range.
Repeat the test using a new test strip as soon as
the measuring environment, measuring device
and test strips have reached room temperature
(+20 °C to +26 °C).
Err
Used or contaminated test strip in-
serted.
•
Insert an unused test strip that has not ex-
pired.
• Repeat the blood glucose measurement.
Err001 System error.
Remove battery and reinsert battery. Contact
Customer Services if the problem persists.
Err002 Insufficient blood on the test strip.
Repeat the measurement using a new test strip.
12
127
130
System error.
Contact Customer Services.
Unknown error messages.
Remove battery and reinsert battery. Contact
Customer Services if the problem persists.