Troubleshooting Guide 

 

 

20

Troubleshooting Guide 

LIFT TROUBLE 

CHECK 

HAND CONTROL DOES NOT RESPOND 

 CHECK LARGE RED EMERGENCY STOP 

BUTTON ON CONTROL BOX 

  CHECK CONNECTOR ON HAND CONTROL 

CORD, 

 CHECK BATTERY CONDITION (REPLACE 

WITH A FULLY CHARGED BATTERY PACK) 

UP & DOWN BUTTONS ON CONTROL BOX DO 
NOT RESPOND. 

 CHECK LARGE RED EMERGENCY STOP 

BUTTON ON CONTROL BOX 

 CHECK BATTERY CONDITION (REPLACE 

WITH A FULLY CHARGED BATTERY PACK) 

ACTUATOR DOES NOT RESPOND 

 CHECK LARGE RED EMERGENCY STOP 

BUTTON ON CONTROL BOX 

 

CHECK BATTERY IS INSTALLED 
CORRECTLY AND FULLY CHARGED. TEST 
WITH A NEW FULLY CHARGED BATTERY 
PACK  

  CHECK HAND CONTROL IS CONNECTED. 

 CHECK ACTUATOR IS CONNECTED TO 

CONTROL BOX. 

AUDIBLE “BEEP” IS HEARD FROM CONTROL 
BOX 

 BATTERY IS LOW, REPLACE WITH A 

FRESHLY CHARGED BATTERY PACK 

ACTUATOR “STALLS” DURING LIFT 

 BATTERY IS LOW, REPLACE WITH A 

FRESHLY CHARGED BATTERY.PACK MAKE 
SURE TO NOT EXCEED THE LIFTING 
CAPACITY. 

CHARGER TROUBLE 

CHECK 

“POWER ON” LIGHT ON CHARGER IS NOT LIT 

  CHECK CHARGER IS PLUGGED INTO WALL 

RECEPTACLE 

CHARGER IS PLUGGED IN BUT “POWER ON” 
LIGHT IS NOT LIT 

 CHECK THAT ELECTRICAL OUTLET HAS 

POWER * 

YELLOW INDICATOR DOES NOT LIGHT WHEN 
BATTERY IS INSERTED IN CHARGER AND THE 
GREEN LIGHT IS “ON” 

 CHECK BATTERY PACK IS PROPERLY 

SEATED IN CHARGE 

BATTERY TROUBLE 

CHECK 

BATTERY PACK IS PROPERLY SEATED BUT 
NO LIGHTS ARE VISIBLE 

  CALL FOR SERVICE (CHARGER MAY BE 

FAULTY) 

YELLOW INDICATOR LIGHT DOES NOT GO 
OFF AFTER SEVERAL HOURS OF CHARGING 
TIME 

 INTERNAL BATTERIES NEED REPLACING 

CALL FOR SERVICE OR CONTACT YOUR 
MAINTENANCE DEPARTMENT. 

BATTERY PACK INDICATES IT’S FULLY 
CHARGED WHEN IN THE CHARGER BUT 
WHEN PLACED IN THE LIFT WILL ONLY DO A 
FEW LIFTS 

  AMPERAGE IN THE BATTERY IS TOO LOW 

AND WILL NOT TAKE A FULL CHARGE, 
REPLACE INTERNAL BATTERY IN THE 
BATTERY PACK SYSTEM** 

 

(*) SOME WALL OUTLETS ARE CONTROLLED BY WALL/LIGHT SWITCHES. ENSURE POWER TO WALL OUTLET 
CONTINUES AFTER WALL/LIGHT SWITCH IS TURNED OFF. 
(**) GENERALLY WITH LOW AMPERAGE THE ACTUATOR WILL EMIT A HUMMING NOISE INDICATING 
INSUFFICENT BATTERY POWER.

Summary of Contents for Ministand

Page 1: ...001 20075 EN rev 6 September 2010 Ministand Instructions for Use...

Page 2: ...e patented or patent pending Our policy is one of continuous development and we therefore reserve the right to make technical alterations without notice The content of this publication may not be copi...

Page 3: ...10 Operating the Ministand 11 Introduction 11 Adjusting the Width of the Legs 11 Power Base Models 12 Correct Use of the Brakes 12 Brakes application and release 12 Emergency Shut Off 12 Battery Pack...

Page 4: ...t to avoid injuries that can be attributed to the use of inadequate parts only parts designated by BHM Medical should be used on equipment and other appliances supplied by BHM Medical Furthermore unau...

Page 5: ...n this manual The Ministand must always be handled by a trained caregiver who must attend the patient during lift application and operation and in accordance with the instructions outlined in this Ins...

Page 6: ...age of the equipment concerned Consumables must be maintained in accordance with published Instructions for Use and the Preventive Maintenance Schedule Equipment Identification The unit s identificati...

Page 7: ...nufacturer This symbol is accompanied by the product s serial number of the manufacturer This symbol indicates separate collection for all batteries as per the WEEE Directive The black bar indicates t...

Page 8: ...atient This mobile patient lift must be operated by a trained attendant Do not attempt to use the lift if you have not been properly trained to do so This patient lift should never be used by patients...

Page 9: ...ry If the maximum load differs between hoist spreader bar and body support unit i e sling then the lowest maximum load must always be used Electric Shock Prevention Read the battery and charger instru...

Page 10: ...Waste of Electronic and Electrical Components as well as local laws and regulations must be followed If not they may explode leak and cause personal injury When returning batteries insulate their ter...

Page 11: ...10 Sling Compatibility For more information about sling use please refer to the documentation that comes with your sling Chest sling 4001C 4001 Total Transfer sling 4000TTC 4000TT Band sling TST X Hy...

Page 12: ...the legs or push it to the far right to open the legs Fig 4 Pedal base The adjustments are made by changing the position of the pedal Press down on the left side of the pedal to close the legs Press d...

Page 13: ...f the unit has to be halted while moving a patient Upon stationing the lift in front of the patient before the patient places his her feet onto the footrest Brakes application and release To apply bra...

Page 14: ...d one second battery pack not included from the wall mounted charging unit not included Fig 11 Figure 11 NOTE to ensure long battery life charge for a minimum of 8 hours before using the lift for the...

Page 15: ...cated on the control box attached to the mast These controls serve two functions They are secondary controls in the event the hand control stops functioning or if two attendants are involved in the tr...

Page 16: ...e sling are correctly installed in the hook Fig 18 If they are not install the loop properly before lifting the patient Figure 18 Ensure the patient s arms are over the sling and that the sling is not...

Page 17: ...in the sling during the transfer 7 Move the patient directly over the point of transfer and lock the wheels on the wheelchair or bed When the patient is positioned against the chair or bed lock the re...

Page 18: ...ith electrical components or internal corrosion If a hot air dryer is used to dry the lift the temperature must not exceed 80 C 176 F Do not use petroleum based solvents or similar since this may dama...

Page 19: ...removed it is imperative that you reinstall all associated safety devices such as cotter pins split rings and lock nuts If one of these devices is defective you must replace it with an original new co...

Page 20: ...en operating the lift Ensure that the footrest is correctly attached to the mast Ensure that the kneepad is in the proper position for the patient Ensure that the kneepad is tightened If the product d...

Page 21: ...TY CHARGER TROUBLE CHECK POWER ON LIGHT ON CHARGER IS NOT LIT CHECK CHARGER IS PLUGGED INTO WALL RECEPTACLE CHARGER IS PLUGGED IN BUT POWER ON LIGHT IS NOT LIT CHECK THAT ELECTRICAL OUTLET HAS POWER Y...

Page 22: ...h IEC 60601 1 WARNING Radio transmitting devices such as mobile telephones two way radios etc should never be used near the Ministand since they can interfere with the function of the lift Cables from...

Page 23: ...Technical Specifications 22 Figure 22...

Page 24: ...r this warranty to be valid the purchaser must present its original proof of purchase at the moment of the claim The defective unit assembly or part must be returned to BHM Medical for inspection The...

Page 25: ...nces shall BHM Medical be liable for consequential incidental or special damages arising in connection with use or inability to use the products In no event shall BHM Medical s liability for breach of...

Page 26: ...001 20075 EN rev 6 September 2010...

Reviews: