30
User Instruction
5. Before programming:
removing the casing cover
Open the battery compartment, as described
in Point 2. open the cover lock using screw-
driver blade no. 1 and then carefully lift up
the cover. The programming socket will now
be visible. Insert the programming cable plug
into the socket.
Note:
Please ensure that the red marking on
the programming cable plug is on the same
Adjustment – programming
your contact mini
(Only for your doctor or hearing aid acoustician)
IMPORTANT INFORMATION:
The contact mini hearing system is available
as a binaural or monaural hearing system.
Your doctor or hearing aid acoustician will
decide which device is worn on the left
and which is worn on the right during the
programming process. In order to label the
device clearly and to attach it to the wearing
system, please assemble the colour rings
supplied onto the electronic casing push
button as follows:
Blue/red ring:
Left/right device as the patient sees.
Note:
The colour rings must be attached, as
these also serve to dampen vibration.