Page 6 Technical Status: 11/2010
Gas Detector
Operating and Installation Instructions
ExDetector HC 150 / HC 150-K
0,4-2V
Br
N
V
Br
4
5
6
Replacing the Sensor
Please note:
Sensors may only be replaced by specialist personnel authorised to do so by
Lang. Do not open the housing when the power supply is switched on!
locknut
seal ring
sensor
Procedure:
●
Switch off the power supply to the detector
●
Remove the housing cover
●
Disconnect the leads from terminals 4, 5 and 6
●
Undo the locknut
●
Unscrew the sensor
●
Screw the replacement sensor in.
Make sure the sealing ring is correctly seated.
●
Screw the locknut on (serrations towards the housing wall)
●
Connect the leads according to their markings
●
Replace the housing cover
●
Switch on the power supply
●
Calibrate the detector
Spare Parts
●
Replacement sensor for ExDetector HC 150 D1
●
Replacement sensor for ExDetector HC 150-K D1