background image

2.   Actuator 

 

ORM0043_2 

Modulating Controller Application Manual 

Page 4 

 

Position Delay

 can be set to delay the actuator moving 

when the required movement is within two times the 
dead zone. This can be used to prevent a noisy command 
signal constantly trying to move the actuator when if the 
signal was stable it should be stationary. If a delay is not 
required then 

Position Delay

 should be set to zero. 

Smaller static dead zones improve accuracy, however 

increases the possibility of the actuator “hunting” for 

position. Smaller moving dead zones mean the actuator 
stops trying to move closer to the required position but 
increases the possibility of actuator overshoot. 

The most accurate method to setup dead zone values is 
to set the static dead zone values to the ideal required 
accuracy and set the moving dead zones to a value 0.2% 
smaller than the static values. Then test actuator 
response by making a number of large and small 
command changes.  

If the actuator is not reaching the required position then 
the moving dead zone values should be reduced. If the 
actuator is overshooting the required position then the 
moving dead zone values should be increased to a 
maximum of 0.1% below the static values. If the actuator 
still overshoots the required position then pulse zones 
can be used as detailed below. 

2.4   Pulse Zones 

Pulse zones can be used to slow down an actuator on 
approach to the required position. This enables the 
actuator to have more controlled positioning and 
reduces the chance of overshoot. An 

Open Pulse Zone

 

and 

Close Pulse Zone

 

can be set to determine if the 

actuator pulses when opening and closing. If pulsing is 
not required then both 

Open Pulse Zone

 

and 

Close Pulse 

Zone

 

should be set to 0%. 

As an example, if the 

Open Pulse Zone

 

is set to 10%, then 

if the command position is 50%, the 

Open Solenoid

 will 

pulse when the actuator position is between 40 and 50%. 

For both open and close directions, four time values can 
be set to determine how the solenoids pulse when within 
the pulse zone. These are 

Open Limit Move

Open Blnce 

Move

Open Limit Pause

 

and 

Open Blnce Pause

 

for 

pulsing the actuator open, and 

Close Limit Move

Close 

Blnce Move

Clse Limit Pause

 and 

Clse Blnce Pause

 

for 

pulsing the actuator closed. 

Move and Pause times can be set for both limits of the 
pulse zone 

 these being the set pulse zone (Limit) and 

when the command and actual position are equal 
(Blnce). A transfer function is then used within the 

application to set the required move and pause pulse 
times depending on the location within the pulse zone. 

The standard way to set pulsing times from Limit to 
Balance is to decrease the Move time and increase the 
Pause time. Doing this means that the solenoid will pulse 
longer and more often further from the balance point, 
and pulse shorter and less often close to the balance 
point therefore slowing the actuator on approach to the 
required actuator position. 

If a stepping actuator is required the pulse zones can be 
used to create this. The 

Open Pulse Zone

 

and 

Close Pulse 

Zone

 

should be set to 100% which means that the 

solenoids will always pulse no matter the position of the 
actuator. By setting all Move time values to the required 
move time, and all Pause time values to the required 
pause time, ensures that the actuator will step equally 
when positioning no matter how far from the required 
actuator position. 

2.5   Actuator Limits 

Actuator limits can be set to determine when the 
actuator is fully open and fully closed.  

The 

Open Limit

 

determines the position at which the 

actuator is considered fully open. If the 

Feedback

 

position is greater than the 

Open Limit

 

for greater than 

the 

Act Limit Delay

 

time then the 

Open Limit

 

output is 

set active. 

The 

Close Limit

 

determines the position at which the 

actuator is considered fully closed. If the 

Feedback

 

position is less than the 

Close Limit

 

for greater than the 

Act Limit Delay

 

time then the 

Close Limit

 

output is set 

active. 

2.6   Hard Seating 

Hard seating can be used to force the actuator open or 
closed when fully open or closed.  

If 

Hard Seat Open

 

is active and 

Hard Seat Fast

 

is off then 

when both the 

Command Signal 

and 

Feedback

 positions 

are above the 

Open Limit

, the 

Open Solenoid

 

will be 

held active forcing the actuator open. If both 

Hard Seat 

Open

 

and 

Hard Seat Fast

 

are active then only the 

Command Signal

 

needs to be above the 

Open Limit

 

to 

hold the 

Open Solenoid

 

active. This can be used to 

ignore pulse zones and fully open the actuator quickly if 
the actuator is normally in a stepping mode. 

If 

Hard Seat Close

 

is active and 

Hard Seat Fast

 

is off then 

when both the 

Command Signal

 

and 

Feedback

 positions 

are below the 

Close Limit

, the 

Close Solenoid

 

will be held 

active forcing the actuator closed. If both 

Hard Seat 

Summary of Contents for Orange EHPC210

Page 1: ...ower Limited The content of this document may not be used for purposes other than that for which it has been supplied and may not be reproduced either wholly or in part nor may it be used by or its co...

Page 2: ...m Values 13 6 4 System Times 14 6 5 HART Variables 14 This manual provides instruction on the setup and analysis of the modulating controller application Section 2 details specific actuator modules in...

Page 3: ...be linked to the input providing the actuator position signal which provides a calibrated 0 100 value In standard operation 4 20mA Cmd Input should be active meaning that the command value is taken f...

Page 4: ...or both limits of the pulse zone these being the set pulse zone Limit and when the command and actual position are equal Blnce A transfer function is then used within the application to set the requir...

Page 5: ...ide When the controller is in automatic mode if the actuator goes into fault the actuator can be forced to drive to a required location If Fault Open and Fault Close are both off or both are set activ...

Page 6: ...n the actuator will position based on the fault override settings High Pressure SP is a high pressure alarm set point which is used to indicate that the pressure has exceeded the maximum allowable acc...

Page 7: ...uator will position based on the fault override settings To reset Overrun Fault Clear Fault must be set active This immediately resets the fault and then Clear Fault is automatically set to off such t...

Page 8: ...settings because it cannot be certain the pressure has not dropped below the minimum allowable supply pressure 4 2 Controller Access Whenever the controller is powered on the Controller On output is s...

Page 9: ...Actuator is fully closed Pump Output Pump is running ACTUATOR POSITION Provides the current actuator position based on the analogue signal If the signal is in fault 0 0 will be shown COMMAND Provides...

Page 10: ...urrently present Below are the statuses that can be displayed Healthy Accumulator is operating within required limits Pump Output Pump is running Acc Press Fault Pressure signal is in fault Pump Fault...

Page 11: ...signal fault Common Fault Set active when there is a signal indicating an external fault Pump On Demand Set active to enable use of pump on demand features Run Pump To Opn Set active to run the pump w...

Page 12: ...mand signal is in fault Common Fault Set active when there is a signal indicating a common fault Pump Output Set active when the pump is required to run Pump Fault Set active when there is a signal in...

Page 13: ...ic DZ Value indicating how much the command signal must be less than the feedback position when the actuator is static before it starts to close Close Moving DZ Value indicating how much the feedback...

Page 14: ...e position when the actuator is opening Close Limit Move Time that the Close Solenoid pulses active when at the pulse zone limit when the actuator is closing Close Blnce Move Time that the Close Solen...

Reviews: