107
DS440IS
9.2
Error messages
Problem
Cause
Remedy
9999999
flashing
The scale is overloaded.
Reduce the weight to a value
below the maximum scale
capacity.
-01-
Converter fault.
Contact Assistance Service.
Scale connector disconnected or
broken.
Switch off the terminal and check
that the connector is properly
connected.
If necessary, disconnect and then
re-connect the connector.
Switch the terminal back on.
-02-
EEPROM error
Switch the terminal off and then
on again.
-04-
RAM checksum error
Switch the terminal off and then
on again.
-05-
PROGRAM checksum error
Switch the terminal off and then
on again.