23
DS440IS
3.2.4
12x0.35 mm² power supplier-terminal connection cable
Ccable/m <250pF/m
Lcable/m<1.5µH/m
Insulation voltage=1000V
Screened cable
3.2.5
6x0.5 mm² cable for the cells, inputs/outputs, analogue output
Ccable<250pF/m
Lcable<0.95µH/m
Insulation voltage=1000V
Screened cable
3.2.6
Optical fibre
200 µm HCD for sections of up to 500m
Mod. TR60 Fort Fibre RS232 transmitter