Independent of the controller operating mode the units of the actual values „
LIGHT kLUX W/m
2
“ or
the doses „
DOSES MLUXh Wh/m
2
“ are displayed down at the left according to the position of the
rocker switch (2).
The actual values of UVA and VIS or their integrated values are displayed accordingly under X.
Every minute the integrated values of UVA and VIS increase by the respective actual value. The dis-
played units are Wh/m
2
and MLUXh. The maximum value of the integrated value display is the respec-
tive value reached with the last addition before exceeding 999.9. The integration display at the MB1
controller will then not increase any more. Recording by APT-COM™ 3 (option, chap. 17.1) can con-
tinue correctly until overflow of the numeral format Floating Point.
With the LUX dose, a change of the smallest display unit 0.1 MLUXh (i.e.100.000 LUXh) takes approx.
25 hours. Therefore it is possible that the target value seems to be already reached at the display, but
no message appears and the tubes are not automatically switched off. This is due to the fact that the
rounded value 0.1 is displayed but not actually reached (possibly only e.g., 0.08 are reached so far).
In Idle Mode and in Program Mode 0.0 is displayed under W, in Manual Mode the last entered target
dose values are displayed under W.
In Manual Mode the illumination equipment is automatically turned on when entering a dose target
value higher than a dose value already reached. Additional manual switching-on can prevent the auto-
matic switch-off when the target dose value has been reached. In Idle Mode and in Program Mode the
illumination equipment can be manually turned off.
Integration continues until the key switch (12) is set to position “OFF”. The integrated values reached
so far continually remain stored but are not displayed. Integration can be continues at any time.
Position RESET during 5 sec. (hold the key): Resetting the integrated values for UVA and VIS together to
zero.
9.2
Toggling between displays of actual and integrated values
When the key switch “LIGHT INTEGRATOR” (12) is in position “ON”, i.e. the integrative function turned
on, the rocker switch (2) located at the lower side of the triangle box permits toggling between the dis-
plays of the actual, or the integrated light values.
•
Rocker switch (2) in Pos. ON (green mark of the rocker switch is visible):
The actual values of UVA and visible light are displayed under X.
The controller shows on its lower left part the units “
LIGHT kLUX W/m
2
”:
Light intensity in the visible spectral range in kLUX, in the UVA range in W / m
2
.
The integration continues in the background.
•
Rocker switch (2) in Pos. OFF (green mark of the rocker switch is not visible):
The integrated values of UVA and visible light are displayed under X.
The controller shows on its lower left part the units “
DOSES MLUXh Wh/m
2
”:
Dose in the visible spectral range in MLUXh, in the UVA range in Wh / m
2
.
Independent of the position of the rocker switch (2) the actual values of UVA and of visible light are dis-
played under X, if:
•
Key switch “LIGHT INTEGRATOR” (12) is set to “OFF”
•
Both set-point values of the target doses have been set to 0.0