KBF-LQC (E5.2) 12/2010
page 49/106
To enter values, select the corresponding line via the
arrow keys.
Hit the ENTER button. The
program editor
appears.
Enter the individual values of the selected program section.
Program editor Abs.Nr. 6
****.*
Setpoint 1
+100.0
FAN
Operating contacts
Time
00:45:00
Repeat Section
5
Repeat Number
10
Tol.-band min.
-1999.0
Tol.-band max.
+9999.0
Parameter set
1
Setpoint 1
Temperature with start of the program section
FAN
Fan speed in %
Operating contacts Operation lines (no function)
Time
Duration of the program section
Repeat Section
No. of start section in case of repeat cycles
Repeat Number
No. of repeats in case of repeat cycles
Tol.-band min.
Tol.-band max.
Temperature limits (maximum / minimum temperature)
(In case of exceeding: temporary program stop)
Parameter set
Pre-selected value (Do NOT change!)
Select the parameters via the arrow keys and confirm by ENTER.
Then enter the values via the arrow keys, and confirm the entry by ENTER.
For a negative set-point entry, enter the numerical value first, and only then the negative sign
(-).
With set-point type “
Limit
”, adapt the safety controller (chap. 11.2) or the temperature safety
device class 3.3 (option, chap. 11.3) to the highest temperature set-point value of the program
actually used. Check the safety device for each temperature program and adapt it if neces-
sary. Set the safety controller set-point or the set-point of temperature safety device class 3.3
(option) by approx. 2 °C to 5 °C above the controller temperature set-point.
Recommended setting: Set-point type “
Offset
” with set-point 2 °C.
Entering the set point values when operating without illumination:
Adapt the setting of the temperature set-point according to the table in chap. 8.2.