KBF-LQC (E5.2) 12/2010
page 52/106
10.5 Program entry as set-point ramp or as set-point step
In order to avoid incorrect programming, we recommend plotting the temperature profile and the humidity
profile (chart templates in chap. 10.11 and 10.12) and entering the values into a table (templates in chap
10.13 and 10.14).
The controller provides 8 operation lines (non-functional with standard unit) that can be activated or de-
activated for each program section.
Program entry as set-point ramp (example of a temperature program)
W/°C
t/min.
01 02
03
04
0
20
40
60
80
100
30 120
180
380
05
Tolerance
maximum
Tolerance
maximum
Program table corresponding to the diagram above:
Program
section
Set-point
temp.
Fan Section
time Operation
lines
Target
section
No. of
cycles
Min.
tolerance
Max.
tolerance
01 40
100
%
00:30:00 Off 1 0 -1999
+9999
02 60
100
%
01:30:00 Off 1 0 -5 +5
03 90
100
%
01:00:00 Off 1 0 -2 +2
04 90
100
%
03:20:00 Off 1 0 -1999
+9999
05 20
100
%
00:00:01 Off 1 0 -1999
+9999
Enter now the values of the above program table into the program table of one of the 25 program places
of the program controller MB1:
08:43:55 15.12.09