KBF-LQC (E5.2) 12/2010
page 73/106
Fluorescent tube cool white:
T8 fluorescent tube in form of a rod with a tube diameter of 26mm. Length
according to chamber size 600 mm or 900 mm. Emissive range in the visible spectral range 400 to 800
nm. The relative spectral distribution meets the F6 standard (cool white) acc. to ISO 10977.
Fluorescent tube “BINDER Q1B Synergy Light”:
T8 fluorescent tube in form of a rod with a tube di-
ameter of 26mm. Length according to chamber size 600 mm or 900 mm. Emissive range in the visible
spectral range 400 to 800 nm. Emissive range in the UVA range 320 to 400 nm.
WARNING
UV light hazard.
Eye and skin injury.
∅
Do NOT look directly into the light.
¾
Minimize skin exposure.
The maximum allowed ultraviolet exposure emitted by this unit on unprotected skin or eye shall not ex-
ceed 7.7 hours per day.
The waste heat of the fluorescent tubes leads to a modification of the temperature - humidity diagram:
When operating the chamber with illumination: Restricted temperature and humidity range 20
°C /
68°F
to 60 °C /
140°F
, at 20 °C not below 30 % r.H.
15.2 Adjustable light cassettes
The KBF-LQC units are equipped with fluorescent tubes for UVA and the visible spectral range. Special
reflector material in the cassettes ensures optimum light diffusion and efficient utilization of the high light
intensity. The lens of the headlight leads to a homogeneous intensity distribution even with short distance
to the shelf. The fluorescent tubes are built in light cassettes that can be freely positioned within wide
areas. They illuminate very homogeneously the racks below them.
KBF-LQC 240
KBF-LQC 720
Figure 25: Positions of light cassettes KBF-LQC