(2) Cable
connections
Connect the plug of the cable to the socket (22) at the rear of the unit.
The socket (22) is marked with a sticker:
Socket (22)
(17)
(18)
Figure 37: Connections at the unit rear KBF-LQC 240
(3) Hose
connections
Plug the freshwater hose to the hose connection (21) above the freshwater can and secure it with a
hose clamp. You can use a part of the regularly delivered water hose.
Screw the hose nozzle (brass) to the free edge of the hose and screw it onto freshwater connection
“IN” (17) at the rear of the unit.
When the freshwater can is empty, 60 seconds later the alarm message HUMID ALARM is displayed on
the controller, the buzzer sounds (chap. 12), and the humidity system turns off.
To guarantee humidification during 24 hours even at high humidity set-points with manual