KBF-LQC (E5.2) 12/2010
page 98/106
Unit size
240
720
Electrical data
IP system of protection acc. to EN 60529
20
20
Nominal voltage (+/-10%) 50/60 Hz
V
200 to 240 / 1N~
Power plug
shock proof plug
Nominal power
kW
2.40
3.50
Energy consumption at 40 °C /
104°F
and 75 % r.H. 4)
Wh/h
600
2350
Installation category acc. to IEC 61010-1
II
II
Pollution degree acc. to IEC 61010-1
2
2
Over-current release category B, 2 poles
Amp
16
16
Noise level approx.
dB (A)
52
53
Electrical connection data KBF-LQC constructed acc. to UL requirements (for the USA und Can-
ada)
Unit size
240
720
Electrical data KBF-LQC-UL
IP system of protection acc. to EN 60529
20
20
Nominal voltage (+/-10%) 60 Hz
V
200 to 240 / 2~
Power plug
NEMA
6-20P
6-20P
Nominal power
kW
2.40
3.50
Installation category acc. to IEC 61010-1
II
II
Pollution degree acc. to IEC 61010-1
2
2
Over-current release category B, 2 poles
Amp
16
16
1)
Lower values are valid at an ambient temperature of max. 25 °C
/ 77°F
2)
to 98 % of the set value
3)
Average value, measured at 25 °C /
77°F
with a spherical sensor (+/-10%) by 12 cm /
4.7 in
below
the light cassette
4)
Use this value for sizing air condition systems
The sensor operating range is up to 60 °C. The temperature controller MB1 has been limited to 60 °C as
soon as at least one of the light sensors is plugged-in. The humidity is irrelevant because the sensor
housing is carried out in IP system of protection IP65.
All technical data is specified for units with standard equipment at an ambient temperature of +25 °C /
77°F
and a power supply voltage fluctuation of +/-10%. The temperature data is determined in accor-
dance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of
10 % of the height, width and depth of the inner chamber. Technical data refers to 100% fan speed.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-
cording to the load.
Bringing a source of humidity into the inner chamber will affect the minimum humidity specifi-
cation and may affect the humidity accuracy.