For surface protection, perform cleaning as quickly as possible.
After cleaning completely remove cleaning agents from the surfaces with a moistened towel.
Soapsuds may contain chlorides and must therefore NOT be used for cleaning.
Decontamination
Disconnect the chamber from the power supply prior to decontamination. Pull the power plug.
You can use the following disinfectants:
Inner chamber
Standard commercial surface disinfectants free from acid or Halogenide.
Alcoholic solutions.
We recommend using disinfectant Art. No. 1002-0022.
In case of impurity of the interior with biological or chemical hazardous goods, there are 3 possible pro-
cedures depending on the type of contamination and of the charging material.
(1) The high performance temperature chamber M can be hot air sterilized at 190 °C for at least 30 min-
utes. All inflammable goods must be removed from the interior before.
(2) Spray the inner chamber with an appropriate disinfectant.
Before start-up, the unit must be absolute dry and ventilated, because explosive gases might form
during the decontamination process.
(3) If necessary, have strongly contaminated inner chamber parts removed by an engineer for cleaning,
or have them exchanged. Sterilize the inner chamber parts in a sterilizer or autoclave.
With every decontamination method, ensure adequate personal safety.
CAUTION
Danger of corrosion.
Damage to the unit.
∅
Do NOT use acidic or chlorine cleaning detergents.
We recommend using the neutral cleaning agent Art. No. Art. Nr. 1002-0016 for a thorough
and mild cleaning.
Any corrosive damage that might arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.
12.3 Sending back the unit to BINDER GmbH
If you send a BINDER product to us for repair or any other reason, we will only accept the product upon
presentation of an authorization number that has previously been issued to you. We will issue an authori-
zation number after receiving your complaint either in writing or by telephone prior to your sending the
BINDER product back to us. The authorization number will be issued following receipt of the information
mentioned below:
•
BINDER product type and serial number
•
Date of purchase
•
Name and address of the dealer from which you bought the BINDER product
•
Exact description of defect or fault
Summary of Contents for APT.line M Series
Page 54: ...M 01 2009 page 54 61 15 6 Dimensions M 53 ENT ENT ENT...
Page 55: ...M 01 2009 page 55 61 15 7 Dimensions M 115 ENT ENT ENT...
Page 56: ...M 01 2009 page 56 61 15 8 Dimensions M 240 6 8 ENT ENT ENT...
Page 57: ...M 01 2009 page 57 61 15 9 Dimensions M 400 ENT ENT ENT...
Page 58: ...M 01 2009 page 58 61 15 10 Dimensions M 720 ENT ENT ENT...