KBF / KBF P (E5.2) 02/2012
page 94/103
Unit size
240
720
Electrical data
IP system of protection acc. to EN 60529
20
20
Nominal voltage (+/-10%)
V
200 to 240 / 1N~
Power frequency
Hz
50/60
(9x20-0160)
50
(9x20-0244)
50/60
(9x20-0143)
50
(9x20-0246)
Power plug
shock proof plug
Nominal power
kW
2.40
3.50
Energy consumption at 40 °C /
104 °F
and 75 % r.H. 4)
Wh/h
600
2350
Installation category acc. to IEC 61010-1
II
II
Pollution degree acc. to IEC 61010-1
2
2
Over-current release category B, 2 poles
Amp
16
16
Noise level approx.
dB (A)
52
53
Electrical connection data KBF P constructed acc. to UL requirements (for the USA and Canada)
Unit size
240
720
Electrical data KBF P-UL
IP system of protection acc. to EN 60529
20
20
Nominal voltage (+/-10%) 50/60 Hz
V
200 to 240 / 2~
Nominal power
NEMA
6-20P
6-20P
Power plug
kW
2.40
3.50
Installation category acc. to IEC 61010-1
II
II
Pollution degree acc. to IEC 61010-1
2
2
Over-current release category B, 2 poles
Amp
16
16
1)
Lower values are valid at an ambient temperature of max. 25 °C /
77 °F
2)
to 98 % of the set value
3)
Average value, measured at 25 °C /
77 °F
with a spherical sensor (±10%) by 12 cm /
4.7 in
below
the light cassette
4)
Use this value for sizing air condition systems
All technical data is specified for unloaded units with standard equipment at an ambient temperature of
+25 °C /
77 °F
and a power supply voltage fluctuation of +/-10%. The temperature data is determined in
accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearanc-
es of 10 % of the height, width and depth of the inner chamber. Technical data refers to 100% fan speed.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-
cording to the load.
Bringing a source of humidity into the inner chamber will affect the minimum humidity specifica-
tion and may affect the humidity accuracy.
Summary of Contents for KBF 115 (E5.2)
Page 5: ...KBF KBF P E5 2 02 2012 page 5 103 Product registration ...
Page 98: ...KBF KBF P E5 2 02 2012 page 98 103 19 7 Dimensions KBF 115 ...
Page 99: ...KBF KBF P E5 2 02 2012 page 99 103 19 8 Dimensions KBF 240 KBF P 240 ...
Page 100: ...KBF KBF P E5 2 02 2012 page 100 103 19 9 Dimensions KBF 720 KBF P 720 ...