19.4 Analog outputs for temperature and humidity (option)
With this option the chamber is equipped with analog outputs 4-20 mA for temperature and humidity.
These outputs allow transmitting data to external data registration systems or devices.
The connection is realized as a DIN socket (3) in the right lateral control panel as follows:
ANALOG OUTPUT 4-20 mA DC
PIN 1: Temperature –
PIN 2: Tempe
PIN 3: Humidity –
PIN 4: Hu
Temperature range: -10 °C /
14 °F
to +100 °C /
212 °F
Humidity range: 0 % r.H. to 100 % r.H.
A suitable DIN plug is enclosed.
Figure 20: Pin allocation of DIN socket (3) for option analog outputs
19.5 Zero-voltage relay alarm outputs for temperature and humidity (option)
The chamber equipment with optional zero-voltage relay outputs for temperature and
humidity (option) permits the transmission of alarms to a central monitoring system.
Connection is established via a DIN socket (6) located on the right lateral control
panel.
Figure 21: Pin configuration of the DIN socket (6)
Temperature contact
Humidity contact
Pin 1: Pin
Pin 2: Make
Pin 3: Pin
Pin 4: Make
In case of a temperature alarm, pins 1 and 2 are open; with humidity alarm, pins 3 and 4 are open. This
happens simultaneously with the alarm message shown on the controller display.
In case of power failure, both contacts are open.
Maximum loading capacity of the switching contacts: 24V AC/DC - 2,5A
DANGER
Electrical hazard.
Danger of death.
Damage to switching contacts and connection socket.
∅
Do NOT exceed the maximum switching load of 24V AC/DC – 2.5A.
∅
Do NOT connect any devices with a higher loading capacity.
A temperature and humidity alarm message will remain visible on the controller display during the whole
time of the alarm transmission via the zero-voltage relay outputs.
As soon as the cause of the alarm is rectified, you can reset the alarm transmission via the zero-voltage
relay outputs together with the alarm message on the controller.
1
2
1
2
Summary of Contents for KBF 115
Page 137: ...KBF KMF E6 07 2017 page 137 155 23 8 Dimensions size 115 ...
Page 138: ...KBF KMF E6 07 2017 page 138 155 23 9 Dimensions size 240 ...
Page 139: ...KBF KMF E6 07 2017 page 139 155 23 10 Dimensions size 720 ...
Page 140: ...KBF KMF E6 07 2017 page 140 155 23 11 Dimensions size 1020 ...
Page 142: ...KBF KMF E6 07 2017 page 142 155 ...
Page 143: ...KBF KMF E6 07 2017 page 143 155 ...
Page 144: ...KBF KMF E6 07 2017 page 144 155 24 2 EU Declaration of Conformity for KMF ...
Page 145: ...KBF KMF E6 07 2017 page 145 155 ...
Page 146: ...KBF KMF E6 07 2017 page 146 155 ...
Page 148: ...KBF KMF E6 07 2017 page 148 155 ...
Page 149: ...KBF KMF E6 07 2017 page 149 155 25 Product registration ...