KBF + KMF (E6) 07/2017
page 24/155
4. Installation and connections
4.1
Spacer for wall distance
Please fix both spacers with the supplied screws at the chamber rear. This serves to ensure the pre-
scribed minimum distance to the rear wall of 100 mm /
3.94
in
.
Figure 8: Spacer for wall distance
Figure 9: Chamber rear with mounted spacers
4.2
Wastewater connection
Fasten the wastewater hose to the wastewater connection “OUT” (14) on the rear of the chamber (olive
∅
14 mm). Observe the following points:
•
You can use a part of the supplied water hose as a drainage hose. In case another hose is used, it
has to be permanently resistant against at least 95 °C /
203 °F
.
•
Mount the wastewater hose with a maximum positive inclination of 1 m and a maximum total length of
3 m.
•
Protect the chamber end of the drainage hose with one of the supplied hose clamps.
•
Reliably prevent sucking back of wastewater. The end of the wastewater hose must not be immersed
in liquids. This can be ensured e.g., by free discharge.
Waste water is collected in an internal can with a volume of approx. 0.5 liters. It is pumped off
only when required, thus there is no continuous waste water flow.
Protect the waste water supply with the supplied hose clamps.
Summary of Contents for KBF 115
Page 137: ...KBF KMF E6 07 2017 page 137 155 23 8 Dimensions size 115 ...
Page 138: ...KBF KMF E6 07 2017 page 138 155 23 9 Dimensions size 240 ...
Page 139: ...KBF KMF E6 07 2017 page 139 155 23 10 Dimensions size 720 ...
Page 140: ...KBF KMF E6 07 2017 page 140 155 23 11 Dimensions size 1020 ...
Page 142: ...KBF KMF E6 07 2017 page 142 155 ...
Page 143: ...KBF KMF E6 07 2017 page 143 155 ...
Page 144: ...KBF KMF E6 07 2017 page 144 155 24 2 EU Declaration of Conformity for KMF ...
Page 145: ...KBF KMF E6 07 2017 page 145 155 ...
Page 146: ...KBF KMF E6 07 2017 page 146 155 ...
Page 148: ...KBF KMF E6 07 2017 page 148 155 ...
Page 149: ...KBF KMF E6 07 2017 page 149 155 25 Product registration ...