KBF + KMF (E6) 07/2017
page 72/155
11.1.2 Alarm messages
Condition
Alarm message
Start after
condition occurred
Zero-voltage relay
alarm output (option)
The current actual temperature value
is outside the tolerance range (chap.
11.4)
“Temperature
range”
after configurable
time
time as alarm start
The current actual humidity value is
outside the tolerance range (chap.
11.4)
“Humidity range
after configurable
time
time as alarm start
Open chamber door
“Door open
after 5 minutes
----
Power failure
---
----
immediately
Exceeded setpoint of the safety control-
ler class 3.1
“Safety controller
immediately
----
Exceeded maximum or minimum tem-
perature (option temperature safety
device class 3.3)
“Temp. safety
device
immediately
----
Temperature sensor defective
e.g. “- - - - “ or
“<-<-< “ or “>->->”
immediately
----
Safety controller temperature sensor
defective
Safety controller
sensor
immediately
----
Alarm messages are displayed in the list of active alarms until acknowledging them. They are also shown
in the event list.
11.1.3 Messages concerning the humidity system
Condition and measures
Message
Start after condition
occurred
The humidity module is defective. Contact BINDER
service
“Humidity system”
immediately
The humidity module cannot fill up.
In case of freshwater supply via water pipe:
The water tap is closed, or the chamber is defective
(e.g. inlet valve of humidity module).
In case of freshwater supply via freshwater can (option,
chap. 19.9):
Water can is empty. Humidification is turned off. In case
of refrigerating operation, the interior is strongly dehu-
midified. When the water supply is functional again, the
humidity system restarts, or the chamber is defective.
“Freshwater supply”
immediately
The humidity module cannot empty the condensate
tank.
Wastewater tube obstructed. Check the length and loca-
tion of the wastewater tube. If appropriate contact
BINDER service.
“Wastewater”
immediately
Purging is required soon. Turn off and on again the
humidity switch to start purging. After a successful purg-
ing the notification resets automatically.
“Humidity system
purging req.”
after predefined time
(approx. 3-5 months
depending on use)
Maintenance of the humidity system is required. Contact
BINDER service.
“Humidity module
service”
after predefined time
(approx. 1 year)
Messages concerning the humidity system are shown in the event list.
Summary of Contents for KBF 115
Page 137: ...KBF KMF E6 07 2017 page 137 155 23 8 Dimensions size 115 ...
Page 138: ...KBF KMF E6 07 2017 page 138 155 23 9 Dimensions size 240 ...
Page 139: ...KBF KMF E6 07 2017 page 139 155 23 10 Dimensions size 720 ...
Page 140: ...KBF KMF E6 07 2017 page 140 155 23 11 Dimensions size 1020 ...
Page 142: ...KBF KMF E6 07 2017 page 142 155 ...
Page 143: ...KBF KMF E6 07 2017 page 143 155 ...
Page 144: ...KBF KMF E6 07 2017 page 144 155 24 2 EU Declaration of Conformity for KMF ...
Page 145: ...KBF KMF E6 07 2017 page 145 155 ...
Page 146: ...KBF KMF E6 07 2017 page 146 155 ...
Page 148: ...KBF KMF E6 07 2017 page 148 155 ...
Page 149: ...KBF KMF E6 07 2017 page 149 155 25 Product registration ...