KBF + KMF (E6) 07/2017
page 108/155
17. Humidification / dehumidification system
The chamber is equipped with a capacitive humidity sensor. This results in a control accuracy of up to +/-
3 % r.H. of the set point. The temperature-humidity diagrams (Figure 19) show the possible working
ranges for humidity.
•
In the “setpoints” menu you can turn humidity control (humidification and dehumidification) on or off
with the setting “Control on/off” (chap. 6.3).
With humidity control turned off, the humidification module cools down. After activation it will take up to
20 minutes until the humidification function is fully available again. This setting is required when
operating the chamber without a water connection in order to avoid humidity alarms.
•
Operation line “Humidity off” serves to turn off the humidification / dehumidification system in Fixed
value operation (chap. 7.3, time program operation (chap. 9.7.3) and week program operation (chap.
10.6.7). This allows configuring the disconnection for individual program sections.
When the humidification / dehumidification system is turned off via operation line it remain on standby
(filled and heated). Therefore it is immediately available after turning on.
The preset temperature and humidity values should be situated within the optimum range
(hatched range in Figure 19).Only within this area will the chamber not be exposed to exces-
sive moisture due to condensation.
In the short-term set points outside the optimum range can also be targeted. The control accu-
racies of
±
3 % r.H., however, cannot be guaranteed in this case.
KMF:
When operating the chamber with activated humidity, humidity control turns off automatically at
temperature set-points below 0 °C /
32 °F
or above 95 °C /
203 °F
. The information icon
Humidity off
is
displayed in the screen header in Normal display. When the temperature setpoint is set back to the range
from 0 °C /
32 °F
to 95 °C /
203 °F
, humidity control turns on again and the information icon
Humidity off
disappears.
0
10
20
30
40
50
60
70
80
90
100
10
20
30
40
50
60
70
80
90
100
% r.F.
°C
A
B
0
10
20
30
40
50
60
70
80
90
100
10
20
30
40
50
60
70
80
90
100
% r.F.
°C
A
B
C
KBF temperature-humidity diagram
KMF temperature-humidity diagram
Figure 19: Temperature-humidity diagrams
Range A:
Control range of temperature and relative humidity, condensation free range
Range B:
Discontinuous range (no continuous operation, up to 24 h)
Range C:
In this range, condensation in the inner chamber is possible
Summary of Contents for KBF 115
Page 137: ...KBF KMF E6 07 2017 page 137 155 23 8 Dimensions size 115 ...
Page 138: ...KBF KMF E6 07 2017 page 138 155 23 9 Dimensions size 240 ...
Page 139: ...KBF KMF E6 07 2017 page 139 155 23 10 Dimensions size 720 ...
Page 140: ...KBF KMF E6 07 2017 page 140 155 23 11 Dimensions size 1020 ...
Page 142: ...KBF KMF E6 07 2017 page 142 155 ...
Page 143: ...KBF KMF E6 07 2017 page 143 155 ...
Page 144: ...KBF KMF E6 07 2017 page 144 155 24 2 EU Declaration of Conformity for KMF ...
Page 145: ...KBF KMF E6 07 2017 page 145 155 ...
Page 146: ...KBF KMF E6 07 2017 page 146 155 ...
Page 148: ...KBF KMF E6 07 2017 page 148 155 ...
Page 149: ...KBF KMF E6 07 2017 page 149 155 25 Product registration ...