KBF + KMF (E6) 07/2017
page 44/155
7.
Set-point entry in “Fixed value” operating mode
In Fixed value operating mode you can enter a temperature set-point, a humidity set-point, the fan speed,
and the switching-state of up to 16 operation lines.
All settings made in Fixed value operating mode remain valid until the next manual change. They are
saved also when turning off the chamber or in case of toggling to Idle Mode or Program Mode.
Setting ranges
Control ranges
Temperature
KBF
-5 °C /
41 °F
up to 70 °C /
158 °F.
0 ºC /
32 °F
up to 70 °C /
158 °F
without humidity
10 °C /
50 °F
up to 70 °C /
158 °F
with humidity
Temperature
KMF
-15 °C /
5 °F
up to 100 °C /
212 °F
-10 °C /
14 °F
up to 100 °C /
212 °F
without humidity
10 °C /
50 °F
up to 90 °C /
194 °F
with humidity
Humidity
KBF
0 % r.H. up to 80 % r.H.
10 % r.H. to 80 % r.H.
see climatic diagrams, chap. 17.
Humidity
KMF
0 % r.H. up to 10 % r.H.
0 % r.H. to 98 % r.H.
see climatic diagrams, chap. 17.
Fan speed
40% up to 100 %
Reduce the fan speed only if required, because the spatial distribution of temperature and
humidity will also be reduced.
Technical data refers to 100% fan speed.
For the control range of temperature and relative humidity, see the temperature / humidity diagrams chap.
17).
With set-point type “
Limit
”, adapt the safety controller (chap. 12.2) or the temperature safety
device class 3.3 (option, chap.12.3) always when you changed the temperature set-point. Set
the safety controller set-point or the set-point of temperature safety device class 3.3 (option)
by approx. 2 °C to 5 °C above the controller temperature set-point.
Recommended setting: Set-point type “
Offset
” with safety controller set-point 2 °C.
When operating without humidity by setting “Control off” (chap. 6.3), the humidity tolerance
range function is automatically deactivated.
When operating without humidity by activated operation line “Humidity off” (chap. 7.3), set the
humidity tolerance range to “0” in order to avoid tolerance range alarms (chap. 11.4).
Summary of Contents for KBF 115
Page 137: ...KBF KMF E6 07 2017 page 137 155 23 8 Dimensions size 115 ...
Page 138: ...KBF KMF E6 07 2017 page 138 155 23 9 Dimensions size 240 ...
Page 139: ...KBF KMF E6 07 2017 page 139 155 23 10 Dimensions size 720 ...
Page 140: ...KBF KMF E6 07 2017 page 140 155 23 11 Dimensions size 1020 ...
Page 142: ...KBF KMF E6 07 2017 page 142 155 ...
Page 143: ...KBF KMF E6 07 2017 page 143 155 ...
Page 144: ...KBF KMF E6 07 2017 page 144 155 24 2 EU Declaration of Conformity for KMF ...
Page 145: ...KBF KMF E6 07 2017 page 145 155 ...
Page 146: ...KBF KMF E6 07 2017 page 146 155 ...
Page 148: ...KBF KMF E6 07 2017 page 148 155 ...
Page 149: ...KBF KMF E6 07 2017 page 149 155 25 Product registration ...