Section 6
The Column Adaptor Plate (CAP)

The Column Adaptor Plate (CAP) interfaces with the Aurum vacuum manifold
to convert the manifold from a plate- to a column-processing system (Figure 6).
The CAP has 18 black luer fittings in a 6 x 3 array and comes supplied with
additional caps. Up to 18 Aurum miniprep columns, 18 Aurum midiprep
columns, or 9 Aurum maxiprep columns* can be accommodated on the CAP
without the need for connectors or other manifold accessories. The CAP will
also accommodate other columns with luer ends.

When vacuum is applied to the manifold, the CAP should self-seat, forming
an airtight seal without the need to press down on the CAP. However, the
application of gentle downward force may occasionally be required to facilitate
seating.

* If more than 6 maxi columns are used, empty the waste collection tray
before the wash step.

Fig. 4.  96-well purification protocol overview.

Bulletin 2667    US/EG    Rev A    01-185    0501

Fig. 5.  Column Adaptor Plate with and without Aurum vacuum

manifold.  

5a:  Blow-up view of luer connection between the Bio-Rad mini and

column adaptor plate, 5b:  CAP with individual columns (maxi’s, midi’s & mini’s).

Aurum Vacuum Manifold: Cat. #732-6470

For more information, call Technical Service 
at 1-800-4BIORAD (1-800-424-6723). Visit us 
on the Web at

www.bio-rad.com/aurum/

Vacuum

manifold

Vacuum

filter flask

Vacuum

regulator

Vacuum

source

Aurum

Vacuum

Manifold: 96-Well
Purification

Protocol Overview

For complete protocol, consult instruction manual.

Tubing provided is 4 ft long and must be cut into 
appropriate pieces before proceeding.

Vacuum Setup Conditions

Vacuum source to manifold

1. 

Use tubing to connect vacuum source to 

right side

of vacuum regulator unit.

2. 

Use tubing to connect 

left side

of vacuum 

regulator to filter flask.

Rubber stopper of vacuum filter flask must 

contain a hole to accommodate serological 

(or comparable) pipet.

Vacuum trap must contain additional sidearm 

for connection to regulator.

3. 

Use tubing to connect filter flask to nozzle 

of manifold top.

Vacuum Purification Conditions

When ready to purify

4. 

Place A stage in center of manifold base.

5. 

Place waste collection tray on top of A stage.

6. 

Place manifold top on base to ensure 

complete contact.

7. 

Place desired Aurum 96-well plate on manifold 

top and proceed with appropriate protocol.

Vacuum Elution Conditions

When ready to elute

8. 

Remove A stage and waste collection tray 

from manifold.

9. 

Place B stage in center of manifold base.

10. 

Place clean 96-well collection plate on top 

of B stage.

11. 

Replace manifold top on base to ensure 

complete contact.

12. 

Proceed with appropriate elution step as 

recommended by protocol.

* The Aurum vacuum manifold and stages should 

be rinsed with water after each use.

a.

b.

Summary of Contents for Aurum

Page 1: ...Instruction Manual For technical service call your local Bio Rad office or in the US call 1 800 4BIORAD 1 800 424 6723 Bio Rad Laboratories 2000 Alfred Nobel Dr Hercules CA USA 94547 510 741 1000 1 800 424 6723 4110115 Rev A ...

Page 2: ...Section 4 The Aurum Vacuum Manifold 3 Section 5 Instrument Setup and Use for 96 well Plates 5 Section 6 The Column Adaptor Plate CAP 7 Section 7 Instrument Setup and Use for the Column Adaptor Plate 8 Section 8 The Vacuum Regulator 11 Section 9 Troubleshooting Guide 13 Section 10 Ordering Information 15 ...

Page 3: ...re fast high quality sample preparation in both formats while maintaining the simplicity of handling associated with vacuum processing The uniquely designed vacuum regulator allows for complete control of negative pressure within the Aurum vacuum manifold Section 2 Kit Components The Aurum vacuum manifold contains the following components Vacuum manifold consisting of manifold top manifold base Co...

Page 4: ... base and top The top has an attached tubular silicone gasket allowing the manifold base and top to seal when a vacuum is introduced Attached to the opening in the manifold top is a flat silicone gasket enabling Aurum 96 well purification plates or the CAP to establish an airtight seal with the manifold top when a vacuum is applied Collectively these two gaskets facilitate self seating of all comp...

Page 5: ... manifold top and place the A stage in the depression in the center of the manifold base The etched letter A should be facing up 7 Place the waste collection tray on top of the A stage Replace the manifold top on the base ensuring complete and uniform contact between the manifold top and base 8 Place the desired 96 well plate on the manifold top and follow the instructions in the protocol for the ...

Page 6: ... us on the Web at www bio rad com aurum Vacuum manifold Vacuum filter flask Vacuum regulator Vacuum source Aurum Vacuum Manifold 96 Well Purification Protocol Overview For complete protocol consult instruction manual Tubing provided is 4 ft long and must be cut into appropriate pieces before proceeding Vacuum Setup Conditions Vacuum source to manifold 1 Use tubing to connect vacuum source to right...

Page 7: ...egative pressure within the manifold Vacuum purification conditions 6 Insert the CAP luer ends up into the depression in the vacuum manifold top Ensure that the CAP rests evenly on the gasket 7 Close the unused luer fittings with the caps provided Close caps by rotating clockwise until light resistance is encountered Excessive tightening of a cap may cause the luer fitting to dislodge when the cap...

Page 8: ... com aurum Neutralized lysate Vacuum manifold Vacuum filter flask Vacuum regulator Vacuum source Elution solution Fig 7 Vacuum regulator with clear representation of the gauge Aurum Vacuum Manifold Column Purification Protocol Overview For complete protocol consult instruction manual Tubing provided is 4 ft long and must be cut into appropriate pieces before proceeding Vacuum Setup Conditions Vacu...

Page 9: ...o Replace vacuum manifold or CAP e g manifold or CAP chips and cracks Purge step in purification If closure of wells or protocol columns e g with sealing tape restores negative pressure no corrective action needed Excessive noise Purification plate or CAP Press down gently on during vacuuming not fully seated plate or CAP to seat Gasket residue preventing Rinse manifold with even contact between d...

Page 10: ...Au ur ru um m p plla as sm miid d 9 96 6 k kiit t 2 x 96 well preps includes 2 grow blocks 2 grow membranes 2 binary units 2 lysate filtration plates and 2 plasmid binding plates 2 collection plates reagents protocol overview and instruction manual Problem Possible Cause Possible Solution Worn manifold gasket Replace gasket Incorrect negative Vacuum source connected Connect vacuum pressure from va...

Reviews: