Section 8
The Vacuum Regulator

The vacuum regulator allows complete control of the negative pressures inside
the Aurum vacuum manifold, allowing optimum levels of negative pressure to
be achieved for efficient sample preparation. This is particularly important
when a gradual increase in negative pressure is required during 96-well plate
purifications. Attached in-line between the vacuum source and the vacuum
filter flask, the regulator measures the difference between ambient pressure
and the pressure within the vacuum manifold in inches of mercury (“ Hg). The
regulator is capable of measuring negative pressures up to -30” Hg (Figure 7).
For other purification kits, please consult the appropriate protocols for
recommended negative pressures.

At various points in the protocols, it is necessary to release the vacuum inside
the manifold so that plates and columns can be removed. To release the
negative pressure, rotate the regulator dial fully counter-clockwise until the
gauge reads < 1” Hg. The vacuum is restored inside the manifold by rotating
the control dial clockwise until the desired negative pressure is attained.

The regulator control dial is located off-center, toward the right side of the
regulator. The vacuum source is attached to the right hose fitting and the
vacuum manifold is attached to the left fitting (Figure 3). 

11

10

Fig. 6.  Column purification protocol overview.

Aurum Vacuum Manifold: Cat. #732-6470

For more information, call Technical Service 
at 1-800-4BIORAD (1-800-424-6723). Visit us 
on the Web at

www.bio-rad.com/aurum/

Neutralized
lysate

Vacuum

manifold

Vacuum

filter flask

Vacuum

regulator

Vacuum

source

Elution
solution

Fig. 7.  Vacuum regulator with clear representation 

of the gauge.

Aurum

Vacuum

Manifold: Column
Purification

Protocol Overview

For complete protocol, consult instruction manual.

Tubing provided is 4 ft long and must be cut into 

appropriate pieces before proceeding.

Vacuum Setup Conditions

Vacuum source to manifold

1. 

Use tubing to connect vacuum source to 

right side

of vacuum regulator unit.

2. 

Use tubing to connect 

left side

of vacuum 

regulator to filter flask.

Rubber stopper of vacuum filter flask must 

contain a hole to accommodate serological 

(or comparable) pipet.

Vacuum trap must contain additional sidearm 

for connection to regulator.

3. 

Use tubing to connect filter flask to nozzle 

of manifold top.

Vacuum Purification Conditions

When ready to purify

4. 

Place A stage in center of manifold base.

5. 

Place waste collection tray on top of A stage.

6. 

Place manifold top on base to ensure 

complete contact.

7. 

Place the column adapter plate (CAP) on 

manifold top.

Ensure luer is end up.

Ensure unused luer fittings are tightly closed.

8.

Place desired Aurum vacuum column(s) into

open luer ends.

9.

Proceed with appropriate purification protocol

for column processing.

Spin Elution Conditions

When ready to elute

10. 

Proceed with appropriate spin elution step as 

recommended by protocol.

* The Aurum vacuum manifold and stages should 

be rinsed with water after each use.

Aurum Vacuum Manifold: Cat. #732-6470

For more information, call Technical Service 
at 1-800-4BIORAD (1-800-424-6723). Visit us 
on the Web at

www.bio-rad.com/aurum/

Bulletin 2671    US/EG    Rev A    01-229    0501

Vacuum

manifold

Vacuum

filter flask

Vacuum

regulator

Vacuum

source

Summary of Contents for Aurum

Page 1: ...Instruction Manual For technical service call your local Bio Rad office or in the US call 1 800 4BIORAD 1 800 424 6723 Bio Rad Laboratories 2000 Alfred Nobel Dr Hercules CA USA 94547 510 741 1000 1 800 424 6723 4110115 Rev A ...

Page 2: ...Section 4 The Aurum Vacuum Manifold 3 Section 5 Instrument Setup and Use for 96 well Plates 5 Section 6 The Column Adaptor Plate CAP 7 Section 7 Instrument Setup and Use for the Column Adaptor Plate 8 Section 8 The Vacuum Regulator 11 Section 9 Troubleshooting Guide 13 Section 10 Ordering Information 15 ...

Page 3: ...re fast high quality sample preparation in both formats while maintaining the simplicity of handling associated with vacuum processing The uniquely designed vacuum regulator allows for complete control of negative pressure within the Aurum vacuum manifold Section 2 Kit Components The Aurum vacuum manifold contains the following components Vacuum manifold consisting of manifold top manifold base Co...

Page 4: ... base and top The top has an attached tubular silicone gasket allowing the manifold base and top to seal when a vacuum is introduced Attached to the opening in the manifold top is a flat silicone gasket enabling Aurum 96 well purification plates or the CAP to establish an airtight seal with the manifold top when a vacuum is applied Collectively these two gaskets facilitate self seating of all comp...

Page 5: ... manifold top and place the A stage in the depression in the center of the manifold base The etched letter A should be facing up 7 Place the waste collection tray on top of the A stage Replace the manifold top on the base ensuring complete and uniform contact between the manifold top and base 8 Place the desired 96 well plate on the manifold top and follow the instructions in the protocol for the ...

Page 6: ... us on the Web at www bio rad com aurum Vacuum manifold Vacuum filter flask Vacuum regulator Vacuum source Aurum Vacuum Manifold 96 Well Purification Protocol Overview For complete protocol consult instruction manual Tubing provided is 4 ft long and must be cut into appropriate pieces before proceeding Vacuum Setup Conditions Vacuum source to manifold 1 Use tubing to connect vacuum source to right...

Page 7: ...egative pressure within the manifold Vacuum purification conditions 6 Insert the CAP luer ends up into the depression in the vacuum manifold top Ensure that the CAP rests evenly on the gasket 7 Close the unused luer fittings with the caps provided Close caps by rotating clockwise until light resistance is encountered Excessive tightening of a cap may cause the luer fitting to dislodge when the cap...

Page 8: ... com aurum Neutralized lysate Vacuum manifold Vacuum filter flask Vacuum regulator Vacuum source Elution solution Fig 7 Vacuum regulator with clear representation of the gauge Aurum Vacuum Manifold Column Purification Protocol Overview For complete protocol consult instruction manual Tubing provided is 4 ft long and must be cut into appropriate pieces before proceeding Vacuum Setup Conditions Vacu...

Page 9: ...o Replace vacuum manifold or CAP e g manifold or CAP chips and cracks Purge step in purification If closure of wells or protocol columns e g with sealing tape restores negative pressure no corrective action needed Excessive noise Purification plate or CAP Press down gently on during vacuuming not fully seated plate or CAP to seat Gasket residue preventing Rinse manifold with even contact between d...

Page 10: ...Au ur ru um m p plla as sm miid d 9 96 6 k kiit t 2 x 96 well preps includes 2 grow blocks 2 grow membranes 2 binary units 2 lysate filtration plates and 2 plasmid binding plates 2 collection plates reagents protocol overview and instruction manual Problem Possible Cause Possible Solution Worn manifold gasket Replace gasket Incorrect negative Vacuum source connected Connect vacuum pressure from va...

Reviews: