background image

IN 45 100 010 • AA • 2 / 12 

Principle 

Ion exchange chromatography (IEX) can be used for purification of biomolecules, such as proteins, peptides and 
oligonucleotides,  by  utilizing  the  difference  in  their  surface  charge.  The  biomolecules  interact  with  the 
immobilized ion exchange groups with opposite charge on the chromatography resin. WorkBeads resins are 
available  with  sulfonate  groups  (WorkBeads  40S),  quaternary  amines  (WorkBeads  40Q)  or  tertiary  amines 
(WorkBeads 40 DEAE) as the ion exchange groups. WorkBeads 40S is a strong cation exchanger and will bind 
positively charged molecules. WorkBeads 40Q and WorkBeads 40 DEAE are strong and weak anion exchangers 
respectively and will bind negatively charged molecules. The structure of the ligands used in WorkBeads 40S, 
WorkBeads 40Q and WorkBeads 40 DEAE are shown in Figure 1. 
 
(A) 

 

 
(B) 

 

 
(C) 

 

Figure 1. Structure of the ligand used in WorkBeads 40S (A), WorkBeads 40Q (B) and WorkBeads 40 DEAE. 

The surface charge of proteins depends on the pH of their environment. When the pH is equal to the isoelectric 
point (pI) of the protein the net charge is zero. At pH values below the pI the net charge will be positive, and at 
a pH above the pI the net charge will be negative. It should be noted that the interaction of the protein depends 
on the presence and distribution of both positive and negative charged groups on the surface (net charge). A 
protein may therefore interact with an ion exchange resin also at the pI. The likelihood of binding to either the 
cation or the anion exchange resin will increase when moving away from the pI.  

Ion exchange chromatography begins with equilibration of the column to establish the desired pH and charging 
the resin with suitable counter ions to the charged ligands on the resin (e.g., the negative sulfonate groups can 
interact with Na

+

 ions and the positive trimethyl amine groups can interact with Cl

-

 ions). It is common to use 

an equilibration buffer composed of a buffer substance to control the pH, and NaCl to include suitable counter 
ions. When the sample is applied, proteins with suitable charge will bind to the charged groups of the resin 
while  displacing  the  counter  ions.  Desorption  of  the  proteins  (elution)  is  carried  out  by  increasing  the 
concentration  of  counter  ion  (salt  gradient  elution).  The  counter  ions  will  displace  the  proteins  as  the  salt 
concentration  increases.  Various  additives  (e.g.,  enzyme  inhibitors,  non-ionic  detergents,  urea  and  low 
concentrations of organic solvent) can be used in samples and buffers for IEX as long as they do not strongly 
interact strongly with the charged groups on the resin or on the protein which could interfere with the protein 
binding to the resin. 

Ion exchange chromatography is one of the most frequently used chromatography techniques because of its 
versatility and ability to separate proteins even with small differences in charge. It is also one of the more cost-
effective chromatography techniques and is therefore excellent for scale-up. 

Summary of Contents for BabyBio DEAE

Page 1: ...r basic proteins i e proteins with a high isoelectric point pI while BabyBio Q and BabyBio DEAE columns are suitable for purification of acidic proteins i e proteins with low pI 1 Choose a suitable pH and buffer for the binding of the target protein One pH unit below pI BabyBio S columns or above pI BabyBio Q and BabyBio DEAE columns is a good starting point 2 Connect the column to the chromatogra...

Page 2: ...therefore interact with an ion exchange resin also at the pI The likelihood of binding to either the cation or the anion exchange resin will increase when moving away from the pI Ion exchange chromatography begins with equilibration of the column to establish the desired pH and charging the resin with suitable counter ions to the charged ligands on the resin e g the negative sulfonate groups can i...

Page 3: ...f a sample that has not been properly clarified may reduce the performance and lifetime of the column The sample should be applied under conditions similar to those of the binding buffer 2 Connect the column Cut off or twist off the end at the outlet of the column see Figure 2 Note It is of high importance to cut off the tip at the very end of the cone preferable using a scalpel Incorrect removal ...

Page 4: ...ate may reduce the yield Applied samples should have a pH that gives the target protein a charge that is opposite the charge of the column resin The pH together with the ionic strength in the sample solution might need adjustment for optimal binding 6 Wash After sample application remove unbound impurities by washing the column with 20 30 CV washing buffer or until desired A280 nm absorbance of th...

Page 5: ...m pressure limit onto the first column If possible the maximum pressure of the chromatography system should be set according to Table 2 Remember always to take the system fluidics contribution to the pressure into account Table 2 Recommended maximum pressure settings for BabyBio columns connected in series Notice that the maximum pressure over each column is always 3 bar No of columns in series Ma...

Page 6: ...fication using BabyBio S BabyBio Q and BabyBio DEAE Other buffers can possible be used Buffer Product Buffer composition Binding buffer BabyBio S 50 mM Na phosphate pH 7 0 Binding buffer BabyBio Q 50 mM Tris HCl pH 7 4 Binding buffer BabyBio DEAE 50 mM Tris HCl pH 7 4 Elution buffer BabyBio S 50 mM Na phosphate 1 M NaCl pH 7 0 Elution buffer BabyBio Q 50 mM Tris HCl 1 M NaCl pH 7 4 Elution buffer ...

Page 7: ... jam and thus reduced diffusion rate into the pores A slightly elevated salt concentration reduces but does not eliminate the interactions with the resin by creating a dynamic adsorption desorption equilibrium that allows further diffusion into the resin thus increasing the binding capacity Tuning the flow rate Flow rate is another factor that can be optimized to improve the binding capacity durin...

Page 8: ...endent on purity and recovery requirements as well as properties of the target protein and the sample Using a gradient elution gives increased purity than step elution but step elution may be necessary to obtain the highest possible concentration of the target protein In order to optimize the salt concentration for step elution an initial gradient test run can be carried out to identify a suitable...

Page 9: ...as ion exchange chromatography This can be carried out quickly and easily using BabyBio Dsalt 1 ml or 5 ml columns see Related products BabyBio Dsalt columns are also a useful alternative to dialysis for larger sample volumes or when samples need to be processed rapidly to avoid degradation Additional purification Ion exchange chromatography is a powerful single protein purification step or combin...

Page 10: ...sin using 1 M NaOH applied by a low reversed flow for 2 hours or overnight is often sufficient CIP of the column can be carried out as followed 1 Wash the column with 5 CV deionized water 2 Apply 3 10 CV of 0 5 1 M NaOH for 15 30 minutes Note The contact time is the important factor treatment with NaOH overnight can be necessary if severely fouled 3 Wash the column with 5 10 CV deionized water or ...

Page 11: ... ml 1 ml 5 ml 1 ml 5 ml Column dimension 7 x 28 mm 1 ml 13 x 38 mm 5 ml 7 x 28 mm 1 ml 13 x 38 mm 5 ml 7 x 28 mm 1 ml 13 x 38 mm 5 ml Recommended flow rate BabyBio 1 ml BabyBio 5 ml 1 ml min 150 cm h 5 ml min 225 cm h 1 ml min 150 cm h 5 ml min 225 cm h 1 ml min 150 cm h 5 ml min 225 cm h Maximum flow rate BabyBio 1 ml BabyBio 5 ml 5 ml min 780 cm h 20 ml min 900 cm h 5 ml min 780 cm h 20 ml min 9...

Page 12: ... ml 1 x 1 ml 2 x 1 ml 5 x 1 ml 10 x 1 ml 45 200 101 45 200 102 45 200 103 45 200 104 BabyBio S 5 ml 1 x 5 ml 2 x 5 ml 5 x 5 ml 10 x 5 ml 45 200 105 45 200 106 45 200 107 45 200 108 BabyBio Q 1 ml 1 x 1 ml 2 x 1 ml 5 x 1 ml 10 x 1 ml 45 100 101 45 100 102 45 100 103 45 100 104 BabyBio Q 5 ml 1 x 5 ml 2 x 5 ml 5 x 5 ml 10 x 5 ml 45 100 105 45 100 106 45 100 107 45 100 108 BabyBio DEAE 1 ml 1 x 1 ml ...

Reviews: