15
1)
Quantitative Method.
Press
,
,
to choose standard sample read (RECORD).
to confirm.
2)
Set Parameter.
According to the menu options, press numeric key to set up fitting method,
wavelength method, units and room, press
to confirm, then press
to enter
wavelength, sample no.,
to confirm.
3)
Sample Test.
Press
to go into quantitative standard sample sheet.
4)
Set Standard Sample.
According to the notice, input the standard sample concentration (Fig.
4-7),
to confirm, press
,
to select the value. After finishing inputting
concentration, press
to enter Abs. test.
5)
Calibrate 100%T/0Abs.
Put the Reference Sample in the light path and press
to
calibrate 100%T/0Abs.
6)
Mark Standard Sample.
Put the sample corresponding with the concentration in the light
path, press
to read. Abs. The value will be displayed on the data sheet, according to
notice to test successively.
7)
Sample Test.
Press
to enter curve, press again to enter quantitative data sheet. Put
the reference in the light path, then press
. Put the sample in the light path, then press
to read value.
Method 2
:
Use Standard Samples (INPUT)
1) Quantitative Method.
Press
,
to choose Standard Sample Input (INPUT),then
press
to confirm.
2) Set Parameter.
According to the menu options, press numeric key to set up fitting method,
wavelength method, units and room, press
to confirm, then press
to enter
wavelength, sample no.,
to confirm.
3) Sample Test.
Press
to enter quantitative standard sample sheet.