29
Fig. 5-4
Step 5, Replace the D2 Lamp
Unscrew the 2 screws on the D2 Flange (No.1 in Fig. 5-5), unplug the connector in the power
board(No. 2)and remove the D2 lamp. Draw on the cotton glove and replace a new lamp. Fix the 2
screws and plug the connector again.
Fig. 5-5
Step 6, Replace W Lamp
Remember the direction of the filament before pull out the W lamp. Be sure that the new
lamp’s filament is in the same direction as before.
Pull out the defected W lamp and draw on the cotton glove. Insert the new W lamp as deep as
possible on the lamp seat. Be sure to keep the filament in the same direction as the old one face.