6. Wiring Diagram

7. Trouble Shooting

Common failures and Solutions:

Phenomena

Causation

Treatment Method

No power supply

a. poor plug contact or line broke

a. Connect the plug and line.

b. Fuse protector is broken.

b. Change the fuse protector.

No temperature

rising inside

container

a. Low setting temperature

a. Readjust and set temperature

b. Heater is broken.

b. Change the heater

c. Temp. controller is broken

c. Change the temperature

controller

d. Temp. sensor is loose.

d. Screw up the sensor nut.

e. Temp. sensor is broken

e. Change the temperature sensor.

No temperature

rising alarm

a. Set temp. of Detached tem.

limiter is low

a. Readjust the temperature 30

above setting temperature.

b. Detached temp. limiter sensor

is broken.

b. Change the detached temperature

limiter sensor

Temperature cannot

reach the setting

point.

a. Exhaust port is fully opened

a. Shut off the exhaust port.

b. The container is overfilled, no

hot air convection.

b. Decrease amount of sample to

improve convection condition.

The fan doesn’t

work.

The fan motor is broken

Stop work and check electric

capacity and motor

Displaying-------

The sensor is broken

Change the sensor

Display STOP

Time-up

Press the program key for 3s to

start

Summary of Contents for BOV-V125F

Page 1: ...Forced Air Drying Oven BOV V30 45 65 125 230 625F User Manual BIOBASE GROUP Version 2020 07...

Page 2: ...are original of the Company original or approved by the Company 3 The related electrical equipment is according to national standards and the use of the manual requirements 4 Product operation is carr...

Page 3: ...tent Preface 1 1 Installation and Operating Conditions 3 2 Product Structure 3 3 Technical Parameters 4 4 Control Panel 5 5 Instructions of Operation 6 6 Wiring Diagram 9 7 Trouble Shooting 9 8 Warran...

Page 4: ...h should be turned off for 5 to 10 minutes depending on the length of the door opening time to reduce the occurrence of temperature and avoid unnecessary loss 1 6 In order to maintain the appearance o...

Page 5: ...3 pcs Power Consumption 800W 1200W 1600W 2300W 3000W 5000W Power Supply 220V 10 50 60Hz 380V 50HZ Internal Size W D H mm 310 310 31 0 350 350 35 0 400 360 45 0 500 450 55 0 600 500 75 0 660 760 125 0...

Page 6: ...trol Panel 1 Measured value display 6 Shift key 2 Setting value display 7 Addend key 3 Alarm 2 indicator 8 Reduction key 4 Alarm 1 indicator 9 Main control output indicator 5 Set key 10 Auto tuning in...

Page 7: ...wer row displays the set temperature Press the SET key shortly the temperature setting state will be entered The parameters can be corrected by the keys press the SET key to enter the time setting sta...

Page 8: ...Integral Time 40s d 0 100 Derivative Time 10s T 1 200 Control Period 20s 2s AL1 20 20 alarm value 5 ET 0 2 Function option the regular time 2 dP 0 1 Point changing 1 PT 1 30min Print gap time 1min Ad...

Page 9: ...or 2 Demonstrate time setting up menu ST If ST is 0 Do not mind time then ET is 1 After the electricity on the instrument Reckon by time beginning ET is 2 While the temperature reaches setting value s...

Page 10: ...Temp sensor is broken e Change the temperature sensor No temperature rising alarm a Set temp of Detached tem limiter is low a Readjust the temperature 30 above setting temperature b Detached temp lim...

Page 11: ...obase would not be liable for any repair of damage caused by improper operation 8 3 If the warranty has been expired Biobase would still responsible for repair with relative charges 8 4 Biobase would...

Page 12: ...inan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd bio...

Reviews: