11
Working environment:
(1) Only is suitable for indoor use;
(2) Altitude not exceeding 2000m;
(3) Ambient temperature: 15℃-35℃;
(4) Relative Humidity: ≤75%;
(5) Atmospheric pressure range: 70 kPa-106 kPa;
(6) Electrical parameters: Consistent with the rated voltage of the Microbiological Safety Cabinet
(See 2.1.5 technical parameter performance index);
(7) Power supply need to be grounded; (Judging method: testing the line phase and the neutral phase
of the power supply with multimeter, the line phase to ground voltage should be grid voltage and
the neutral phase to ground voltage should be 0, otherwise the power supply ground is bad).
(8) Overvoltage category: II
(9) Pollution degree: 2
3.4 Installation
a. Remove all the package materials;
b. Inspect the surface of main body to make sure whether there is scratch, deformation or
uncorrelated things;
c. Check the attachment and information according to the packaging list in instruction.
d. Move the whole device to the final installation location, where is convenient to install.
e. This equipment shouldn’t be hosting. Wooden packaging should be used as tap support in
transport process, carried by hydraulic truck or forklift.
f. The whole power line 3 * 1.0mm² * 3000mm; the whole ground line is yellow-green Φ1.0mm2.
The base stand will be packed at back of main body, please take it out before installation.
DO NOT INVERT, DISASSEMBLE OR TITILE THE CABINET during transportation.
g. The base stand assembly
Referring to Picture 4 assemble the base stand.
Note: The base mounting bolts are fixed to the base stand. When install the base stand,
except the 4 bolts marked with red circle in Picture 4, please remove the other bolts and install.
Picture4
M10*20 Hexagon
socket head screws
The bolts in the red ring
shall not be removed