background image

17

9. Alarm status

23. Filter working time

10. Mute status

24. UV lamp working time

11. Filter changing status

25. Reservation timing

12. System working time
13. Glass window down status
14. Glass window up status

a) LCD Screen
The working status of the equipment and operation can be seen on the LCD screen.

b) Soft touch button.
BSC’s main functions could be executed by touch-buttons. User can operate the BSC either by
pressing the buttons on control panel or using the remote control. There are totally 7 common button
on control panel.

: The power button

: To control fluorescent lamp

: To control UV lamp.(It works only after front window fully closed.)

: To control blower working status. (It will not work when front window is fully

closed.)

: Press MUTE button to stop voice prompt

: Press UP button, glass window will raise.

: Press Down button, glass window will fall down.

2.4 Remote Control

It is inconvenient for the users to operate from a distance. Small & light remote control is flexibly to
be used to control all the functions of the cabinet in a distance ≤6m, 30°. The operator can even carry
it with themselves during experiment for convenience.

Summary of Contents for BSC-4FA2

Page 1: ...Microbiological Safety Cabinet BSC 4FA2 4 USER MANUAL BIOBASE GROUP Version 2020 07...

Page 2: ...r damage or for any direct or indirect damage that may occur during the use of the equipment 1 Malfunction or damage due to violation of the instructions precautions and intended use of this manual 2...

Page 3: ...structions 11 2 1 Functions 11 2 1 1 Product Concept 11 2 1 2 Application Range 11 2 1 3 Working theory Air flow pattern and protected area 11 2 1 4 Protected objects 12 2 1 5 TECHNICAL PARAMETERS 12...

Page 4: ...ce period 23 2 6 4 Maintenance methods 23 2 6 5 Storage conditions 23 2 7 Methods and procedures for disinfection 24 2 8 Replacement parts list 26 2 9 Wiring diagram 28 III Trouble shooting and Labels...

Page 5: ...firstly check if packing box is in good condition If the packing box is damaged please take photos 1 1 Unpacking Choose the proper unpacking method according to the actual situation For wooden box 1...

Page 6: ...ecessary tools for unpacking Electric drill with hexagon dead M8 Picture 2 Rapid unpacking diagram Picture 3 Disassemble the screws shown in the below Picture then move the wooden pieces to right and...

Page 7: ...ding battery 1pc Power Line 1pc User manual 1pc Test report 1pc Quality certification card 1pc Drain Valve 1set Stainless steel hex cylinder head bolt 5pcs Stainless steel flat washer10 5pcs Stainless...

Page 8: ...e Humidity 75 5 Atmospheric pressure range 70 kPa 106 kPa 6 Electrical parameters Consistent with the rated voltage of the Microbiological Safety Cabinet See 2 1 5 technical parameter performance inde...

Page 9: ...in Picture 5 please remove the other bolts and install Picture 4 Remove the M10 20 Hexagon socket head screws Buckle plug M10 Stainless steel flat washer M10 Stainless steel spring washer M10 Stainles...

Page 10: ...10 from the accessory box and fasten tightly according to Picture 6 h Adjustment of Footmaster Caster Picture 7 Clockwise rotate caster red part to low down the base feet and the height of the cabine...

Page 11: ...ske Inner diameter outer diameter thickness 13 19 2mm t Drain valve assembling from up to down as Picture 8 illustrated Don t place equipment where dificult to operate and diconnect device 1 5 Checkin...

Page 12: ...the polluted air flow become pure after processed by HEPA ULPA filter When it s used in microbiology experiment environment filled with volatile or toxic chemical and radionuclide suitable exhaust hoo...

Page 13: ...e W D H 1210 600 660mm Total Airflow Volume 465m3 h Downflow Velocity 0 33m s Inflow Velocity 0 53m s Fluorescent lamp Consumption 28W 2 UV Lamp Consumption 30W HEPA Filter 99 999 Diameter 0 3 m Noise...

Page 14: ...n and detect the HEPA filter the leakage rate at any point should not be 0 01 5 Vibration amplitude The net vibration amplitude between frequency 20Hz and 2000KHz is no more than 5 m rms 6 Illuminatio...

Page 15: ...consists of front window hauling sash and position switch 2 Air Filtration System Air Filtration System is the most important system of BSC It consists of blower supply filter and exhaust filter The...

Page 16: ...ich can be controlled by SOCKET button 1 Please make sure the total load of sockets should be 230V 1 5A 2 Fuse protector see Picture 13 The equipment is equipped with main power fuse waterproof socket...

Page 17: ...eel with anti powder coating Strong and steady c Work area is fully made of 304 stainless steel which looks beautiful and with corrosion resistance performance d Base stand is made of cold rolled stee...

Page 18: ...remote control There are totally 7 common button on control panel The power button To control fluorescent lamp To control UV lamp It works only after front window fully closed To control blower worki...

Page 19: ...onfirm and then adjust the minutes and hours with the same way c After the time is confirmed then press related button following which the icon in display lighting d Press the POWER button the reserva...

Page 20: ...aced in parallel Samples should be placed in the cabinet parallel to avoid cross contamination between samples and blocking back air grille 5 In order to avoid samples being sucked into the negative p...

Page 21: ...ring requirements ground bearing capacity 450KPa 17 Total weight 220Kg 18 Allowable pressure of water socket 0 8MPa Optional 19 For experiments that produce waste liquid please timely drain waste wate...

Page 22: ...t conditions it should be replace when the UV timer on the display indicate the working time reaches to 1000 hours e Please move the front window at 200mm height from the work table turn on the fan ma...

Page 23: ...hair brush cleaning twice then wipe and wash with water or other materials waste liquid discharges from the sump drain valve in the left front then clean the sump Before fumigation should not open th...

Page 24: ...rface of operating area excluding the wind distributing grid of operating area and the inner surface of glass door with 70 medical alcohol or household bleach diluted 1 100 i e 0 05 sodium hypochlorit...

Page 25: ...and the time required for disinfection There are some cases need to use these alternative methods e g slow virus disease The disinfection methods should be negotiated by the users and the authenticati...

Page 26: ...re removing the aluminum foil at the start in neutralization stage Removing the aluminum foil does not allow leaking out formaldehyde from Microbiological Safety Cabinet Example 2 The safety cabinet w...

Page 27: ...h preferably overnight q Prepar according to step g power the heating device containing ammonium bicarbonate and Microbiological Safety Cabinet to make ammonium bicarbonate vanish Like operating with...

Page 28: ...W 7 Upper filter Exhaust filter 800 430 69 8 Downflow filter Supply filter 1240 500 69 9 Fan DH230A2 AG5 01 10 Control panel LCD control board strong circuit board weak circuit board display screen 11...

Page 29: ...28 2 9 Wiring diagram BSC 4FA2 4 Wiring diagram Picture 13...

Page 30: ...t setting value is 200mm 2 HEPA filter pressure difference alarm There will be audio and visual alarm if pressure of air supply filter or exhaust filter can t meet present value at the same time LCD d...

Page 31: ...the front window fluorescent lamp and the blower is open or not Control panel Change it Button doesn t work Control panel Make sure the power connects well and the fuse is well Check if the button is...

Page 32: ...trical parts must be operated by a qualified electrician in safety conditions cutting off power supply The other parts are not allowed to remove otherwise the user should take responsibility by them 2...

Page 33: ...32 Picture 14...

Page 34: ...e correspondence type of lamp and put it to the lamp holder and and screw 90 in reverse direction After replacing the UV lamp it needs to keep pressing the button of UV for about five seconds when the...

Page 35: ...bel Description 3 2 1 Company label 3 2 2 Biological hazard label Picture 18 Picture 19 3 2 3 Power plug fuse Picture 20 3 2 4 Ground label Picture 21 3 2 5 Glass door super elevation warning label Pi...

Page 36: ...After the expiration of warranty our company is also responsible for repairs but the corresponding maintenance cost should be charged 4 Life time of Microbiological Safety Cabinet is 8 years from pro...

Page 37: ...inan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd bio...

Reviews: