background image

Note :

• As the forehead measurement temperature is likely to be affected by sweat, oil 

  and the surroundings, the reading shall be taken as a reference only.

• If the detecting sensor is placed at an angle close to the forehead measurement, 

  the reading will be affected by surrounding temperature. Babies’ skin reacts very 

   quickly to the ambient temperature. Therefore, do not take their temperature with

   MINI during or after breastfeeding because the skin temperature may then be 

lower than the internal body temperature.

• If the reading is < 32°C (89.6°F), the display will show "L" together with red 

   backlight and three short beeps.

• If the reading is ≥ 32°C (89.6°F) and < 37.5 °C (99.5°F), the display will show the

   reading together with green backlight and three short beeps.

• If the reading is ≥ 37.5 °C (99.5°F) and < 43°C (109.4°F), the display will show the 

  reading together with red backlight and ten short beeps.

• If the reading is ≥ 43°C (109.4°F), the display will show "H" together with red 

  backlight and ten short beeps.

• The interval of each next measurement is at least 3 seconds.

• The device remains in standby mode for another measurement with last readings 

  stays on the display, °C or °F is blinking. The device cannot perform another 

   measurement while the previous measurement still appears on the display with 

   backlight that lasts for 5 seconds.

• The thermometer will automatically turn off if it is left idle for 25 seconds.

Measuring an Object Temperature

1. Press the On/Scan button for 2~5 seconds when the thermometer is off to enter 

    into object mode. (f) The full display will be shown on the screen for 2 seconds 

    and then the last memory will appear for another 2 seconds. (j)

2. When °C or °F blinks (d), point the torchlight to the object. (k) Make sure the 

    detecting sensor is flat and close to the object, not at an angle.

Note : Perform an object measurement with a distance within 3 cm.

3. Press and release the On/Scan Button, the device will take measurement.

4. Read the result. (l)

The measurement result will be displayed once releasing the On/Scan Button. The 

reading is shown together with green backlight and three short beeps.

Note : 

• If the reading is < 0°C (32°F), the display will show "L" together with red backlight 

  and three short beeps.

• If the reading is ≥ 0°C (32°F) and < 100 °C (212°F), the display will show the reading 

   together with green backlight and three short beeps.

• If the reading is ≥ 100°C (212°F), the display will show "H" together with red 

  backlight and ten short beeps.

• The interval for the next measurement should be 3 seconds.

• The device remains in standby mode for another measurement with last readings 

  stays on the display, °C or °F is blinking. The device cannot perform another 

  measurement while the previous measurement still appears on the display with 

  backlight that lasts for 5 seconds.

• The thermometer will automatically turn off if it is left idle for 25 seconds.

Recalling the Memory

Your thermometer stores 20 most recent body temperature readings.

1. Press the On/Scan button for 5~8 seconds to enter the memory mode. (f)

    Three dashes ("---") will appear with blinking "        " and greenbacklight. (m) 

    Be sure the thermometer is OFF before entering the memory mode.

2. Press the On/Scan Button repeatedly to review the results stored in the device. 

     Backlight in green or red will appear according to the memory reading. When the 

    last record appears on the screen, press the On/Scan Button again to return to the 

Mini Owner’s Manual

first 

 record.

Note :

• Only the body temperature records can be read in the memory mode.

• If there is no readings existed, the "---" icon (with "      " icon blinks) will be shown

  on the screen.

• When the memory is full, the oldest result will be overwritten.

Appendix

Maintenance
Replacing the Battery

The thermometer comes with one 3V CR2032 lithium battery.

When the power is getting low, the "             " icon appears on the screen.

To change the battery, do the following :

1. Press the buckle on battery cover and lift up to remove the cover. (n)

2. Remove the old battery with battery hook by lifting the battery slightly upward 

    and replace with one new 3V CR2032 lithium battery. (o)

3. Close the battery cover.

CAUTION

RISK OF EXPLOSION IF BATTERY IS REPLACED BY AN

INCORRECT TYPE.

DISPOSE OF USED BATTERIES ACCORDING TO THE

INSTRUCTIONS.

Note :

• Although the thermometer works when the "          " icon appears, we still      

  recommend that you change the battery to obtain an accurate result.

• Keep away the battery from small children. If swallowed, promptly seek medical 

   assistance.

• Battery may leak chemicals if unused for a long time. Remove the battery if you

  are not going to use the device for an extended period.

• Properly dispose of the used battery according to your local environmental 

   regulations.

Caring for Your Device

Cleaning

• The detecting sensor is not waterproof. Please wipe with a clean and dry cotton

   swab to clean the detecting sensor on the inside.

• Use a soft and dry cloth to clean the thermometer. Do not use abrasive cleaners.

   Storage.

• The body of the thermometer is not water-resistant. Never put the thermometer 

   under a running tap or submerge it into water.

• Store the thermometer in a cool and dry location. Keep free from dust and away 

   from direct sunlight.

Normal Body Temperature & Fever

Information

The temperature in the forehead and temple area differs from the internal 

temperature, which is taken orally or rectally.

Vasoconstriction, an effect which constricts the blood vessels and cools the skin, 

can occur during the early stages of a fever. In this case, the temperature measured 

by the Biocare MINI thermometer may be unusually low.

If the measurement therefore does not match the patient’s own perception or is 

Mini Owner’s Manual

unusually low, repeat the measurement every 15 minutes. As a reference, you can 

also measure the internal body temperature using a conventional oral or rectal

thermometer.

Body temperature can vary from one individual/person to next. It also varies by 

location on the body and time of day. Below shows the statistical normal ranges 

from different sites. Please keep in mind that temperatures measured from 

different sites, even at the same time, should not be directly compared. Fever

indicates that the body temperature is higher than normal.

This symptom may be caused by infection, overdressing or immunization. Some 

people may not experience fever even when they are ill. These include, but are 

not limited to, infants younger than 3 months old, persons with compromised

immune systems, persons taking antibiotics, steroids or antipyretics (aspirin, 

ibuprofen, acetaminophen), or persons with certain chronic illnesses. Please 

consult your physician when you feel ill even if you do not have fever.

Table*

1

 Body Site Normal Temperature Range

*

1

. Body Temperature at WebMD;

website: http://firstaid.webmd.com/body-temperature retrieved at 2010 Jan 7.

System Troubleshooting

If you follow the recommended action but the problem persists, or error 

messages other than the ones below appear, please call your local customer 

service. Do not attempt to repair yourself and never try to disassemble the device 

under any circumstances.

Measurement Result

Body Site

Temperature Range

Oral

0.6°C (1°F) or more above or below 37°C (98.6°F)

Rectal/ear 0.3°C to 0.6°C (0.5°F to 1°F) higher than oral 

temperature

Axillary 
(armpit)

0.3°C to 0.6°C (0.5°F to 1°F) lower than oral 
temperature

Message

What It Means

Measured temperature falls outside the 
displayed temperature range of 32°C to 43°C 
(89.6°F to 109.4°F), is shown on the display 
with red backlight.

• 

Hi-temperature 

 43°C (109.4°F)

• 

Low-temperature < 32°C (89.6°F)

Measured temperature falls in normal 
reading, within the range of 32°C to 37.5°C 
(89.6°F to 99.5°F), is shown on the display 
with green backlight.

Measured temperature falls within the range 
of 

 37.5°C (99.5°F) and < 43°C (109.4°F), is 

shown on the display with red backlight.

 Mini Owner’s Manual

Message

Cause

What To Do

The ambient temperature 
is below system operation 
range, or the object 
temperature is below 
displayed temperature range.

Place the 
thermometer 
under operating 
temperature range 
of 16°C to 40°C 
(60.8°F to 104°F), 
or under object 
temperature range 
of 0°C to 100°C 
(32°F to 212°F).

The ambient temperature 
is above system operation 
range, or the object 
temperature is above 
displayed temperature 
range.

The battery cannot provide 
enough power for a test.

Replace 
the battery 
immediately.

The thermometer is broken 
or affected by EMF.

Please contact 
customer service.

Troubleshooting

Symbol Information

Symbol Referent

Read instructions before use

Manufacturer

Serial number

Authorized representative in the European 
Community

Type BF Equipment

Collection for electrical and electronic equipment

Mini Owner’s Manual

Specifications

Memory

20 measurement results

Dimensions

100 x 30.5 x 18 mm

Weight

32 g (include battery)

Power Source

One 3V CR2032 lithium battery

Displayed 
temperature range

Forehead: 32°C to 43°C (89.6°F to 
109.4°F)
Object: 0°C to 100°C (32°F to 212°F)

Display resolution

0.1°F / 0.1°C

Accuracy

Body mode :

• 

±0.2°C (±0.36°F) from 36.0°C (96.8°F) 
to 39.0°C (102.2°F)

• 

±0.3°C (±0.54°F) from 32.0°C (89.6°F) 
to 35.9°C (96.6°F) and from 39.1°C 
(102.4°F) to 43.0°C ( 109.4°F)

Object mode :

• 

±1°C (±2°F) from 0°C (32°F) to 100°C 
(212°F)

Temperature unit

°C or °F

Operating 
temperature range

16°C to 40°C (60.8°F to 104°F)

Operating humidity

95% RH or less

Storage 
temperature range

-20°C to 50°C (-4°F to 122°F)

Storage humidity

95% RH or less

Mini Owner’s Manual

Device Standard:

Device Corresponds to the requirements of the standard for infrared 

thermometers :

• ASTM E1965-98

 • EN 60601-1-4 :1996

• EN 12470-5 

• EN 60601-1-6:2010

• EN 60601-1-2:2007/AC:2010 • EN 60601-1

Electromagnetic Compatibility :

Device fulfills the stipulations of the standard EN 60601-1-2.

The stipulations of EU-Directive 93/42/EEC for Medical Devices Class Ila have been 

fulfilled.

WARRANTY TERMS AND CONDITIONS

With respect to disposable products, Biocare warrants to the original purchaser 

that, at time of delivery, each standard product manufactured by Biocare shall be 

free of defects in material and workmanship and, when used for the purposes 

and indications described on the labeling, is fit for the purposes and indications 

described on the labeling. All warranties for a product shall expire as of the 

product expiration date, or if none, after 14 months from the date of shipment 

from Biocare  . Biocare warranty hereunder shall not apply if: (i) a product is not 

used in accordance with its instructions or if it is used for a purpose not indicated 

on the labeling; (ii) any repairs, alterations or other work has been performed by 

Buyer or others on such item, other than work performed with Biocare ’s 

authorisation and according to its approved procedures;  or (iii) the alleged defect 

is a result of abuse, misuse, improper maintenance, accident or the negligence of 

any party other than Biocare  . The warranty set forth herein is conditioned upon 

proper storage, installation, use and maintenance in accordance with applicable 

written recommendations of Biocare . 

0197

Shanghai International Trading Corp. 
GmbH(Hamburg)
Eiffestrasse 80, 20537 Hamburg, Germany
TEL:+49‐40‐2513175
FAX:+49‐40‐255726

Biocare Asia Co., LTD
No.260,Mayun Road, New District Suzhou,
Jiangsu,215129,P.R.C
TEL:+86‐512‐68091772  
FAX:+86‐512‐68097761 
www.bca-medical.com

Reviews: