Note :
• As the forehead measurement temperature is likely to be affected by sweat, oil
and the surroundings, the reading shall be taken as a reference only.
• If the detecting sensor is placed at an angle close to the forehead measurement,
the reading will be affected by surrounding temperature. Babies’ skin reacts very
quickly to the ambient temperature. Therefore, do not take their temperature with
MINI during or after breastfeeding because the skin temperature may then be
lower than the internal body temperature.
• If the reading is < 32°C (89.6°F), the display will show "L" together with red
backlight and three short beeps.
• If the reading is ≥ 32°C (89.6°F) and < 37.5 °C (99.5°F), the display will show the
reading together with green backlight and three short beeps.
• If the reading is ≥ 37.5 °C (99.5°F) and < 43°C (109.4°F), the display will show the
reading together with red backlight and ten short beeps.
• If the reading is ≥ 43°C (109.4°F), the display will show "H" together with red
backlight and ten short beeps.
• The interval of each next measurement is at least 3 seconds.
• The device remains in standby mode for another measurement with last readings
stays on the display, °C or °F is blinking. The device cannot perform another
measurement while the previous measurement still appears on the display with
backlight that lasts for 5 seconds.
• The thermometer will automatically turn off if it is left idle for 25 seconds.
Measuring an Object Temperature
1. Press the On/Scan button for 2~5 seconds when the thermometer is off to enter
into object mode. (f) The full display will be shown on the screen for 2 seconds
and then the last memory will appear for another 2 seconds. (j)
2. When °C or °F blinks (d), point the torchlight to the object. (k) Make sure the
detecting sensor is flat and close to the object, not at an angle.
Note : Perform an object measurement with a distance within 3 cm.
3. Press and release the On/Scan Button, the device will take measurement.
4. Read the result. (l)
The measurement result will be displayed once releasing the On/Scan Button. The
reading is shown together with green backlight and three short beeps.
Note :
• If the reading is < 0°C (32°F), the display will show "L" together with red backlight
and three short beeps.
• If the reading is ≥ 0°C (32°F) and < 100 °C (212°F), the display will show the reading
together with green backlight and three short beeps.
• If the reading is ≥ 100°C (212°F), the display will show "H" together with red
backlight and ten short beeps.
• The interval for the next measurement should be 3 seconds.
• The device remains in standby mode for another measurement with last readings
stays on the display, °C or °F is blinking. The device cannot perform another
measurement while the previous measurement still appears on the display with
backlight that lasts for 5 seconds.
• The thermometer will automatically turn off if it is left idle for 25 seconds.
Recalling the Memory
Your thermometer stores 20 most recent body temperature readings.
1. Press the On/Scan button for 5~8 seconds to enter the memory mode. (f)
Three dashes ("---") will appear with blinking " " and greenbacklight. (m)
Be sure the thermometer is OFF before entering the memory mode.
2. Press the On/Scan Button repeatedly to review the results stored in the device.
Backlight in green or red will appear according to the memory reading. When the
last record appears on the screen, press the On/Scan Button again to return to the
Mini Owner’s Manual
first
record.
Note :
• Only the body temperature records can be read in the memory mode.
• If there is no readings existed, the "---" icon (with " " icon blinks) will be shown
on the screen.
• When the memory is full, the oldest result will be overwritten.
Appendix
Maintenance
Replacing the Battery
The thermometer comes with one 3V CR2032 lithium battery.
When the power is getting low, the " " icon appears on the screen.
To change the battery, do the following :
1. Press the buckle on battery cover and lift up to remove the cover. (n)
2. Remove the old battery with battery hook by lifting the battery slightly upward
and replace with one new 3V CR2032 lithium battery. (o)
3. Close the battery cover.
CAUTION
RISK OF EXPLOSION IF BATTERY IS REPLACED BY AN
INCORRECT TYPE.
DISPOSE OF USED BATTERIES ACCORDING TO THE
INSTRUCTIONS.
Note :
• Although the thermometer works when the " " icon appears, we still
recommend that you change the battery to obtain an accurate result.
• Keep away the battery from small children. If swallowed, promptly seek medical
assistance.
• Battery may leak chemicals if unused for a long time. Remove the battery if you
are not going to use the device for an extended period.
• Properly dispose of the used battery according to your local environmental
regulations.
Caring for Your Device
Cleaning
• The detecting sensor is not waterproof. Please wipe with a clean and dry cotton
swab to clean the detecting sensor on the inside.
• Use a soft and dry cloth to clean the thermometer. Do not use abrasive cleaners.
Storage.
• The body of the thermometer is not water-resistant. Never put the thermometer
under a running tap or submerge it into water.
• Store the thermometer in a cool and dry location. Keep free from dust and away
from direct sunlight.
Normal Body Temperature & Fever
Information
The temperature in the forehead and temple area differs from the internal
temperature, which is taken orally or rectally.
Vasoconstriction, an effect which constricts the blood vessels and cools the skin,
can occur during the early stages of a fever. In this case, the temperature measured
by the Biocare MINI thermometer may be unusually low.
If the measurement therefore does not match the patient’s own perception or is
Mini Owner’s Manual
unusually low, repeat the measurement every 15 minutes. As a reference, you can
also measure the internal body temperature using a conventional oral or rectal
thermometer.
Body temperature can vary from one individual/person to next. It also varies by
location on the body and time of day. Below shows the statistical normal ranges
from different sites. Please keep in mind that temperatures measured from
different sites, even at the same time, should not be directly compared. Fever
indicates that the body temperature is higher than normal.
This symptom may be caused by infection, overdressing or immunization. Some
people may not experience fever even when they are ill. These include, but are
not limited to, infants younger than 3 months old, persons with compromised
immune systems, persons taking antibiotics, steroids or antipyretics (aspirin,
ibuprofen, acetaminophen), or persons with certain chronic illnesses. Please
consult your physician when you feel ill even if you do not have fever.
Table*
1
Body Site Normal Temperature Range
*
1
. Body Temperature at WebMD;
website: http://firstaid.webmd.com/body-temperature retrieved at 2010 Jan 7.
System Troubleshooting
If you follow the recommended action but the problem persists, or error
messages other than the ones below appear, please call your local customer
service. Do not attempt to repair yourself and never try to disassemble the device
under any circumstances.
Measurement Result
Body Site
Temperature Range
Oral
0.6°C (1°F) or more above or below 37°C (98.6°F)
Rectal/ear 0.3°C to 0.6°C (0.5°F to 1°F) higher than oral
temperature
Axillary
(armpit)
0.3°C to 0.6°C (0.5°F to 1°F) lower than oral
temperature
Message
What It Means
Measured temperature falls outside the
displayed temperature range of 32°C to 43°C
(89.6°F to 109.4°F), is shown on the display
with red backlight.
•
Hi-temperature
≥
43°C (109.4°F)
•
Low-temperature < 32°C (89.6°F)
Measured temperature falls in normal
reading, within the range of 32°C to 37.5°C
(89.6°F to 99.5°F), is shown on the display
with green backlight.
Measured temperature falls within the range
of
≥
37.5°C (99.5°F) and < 43°C (109.4°F), is
shown on the display with red backlight.
Mini Owner’s Manual
Message
Cause
What To Do
The ambient temperature
is below system operation
range, or the object
temperature is below
displayed temperature range.
Place the
thermometer
under operating
temperature range
of 16°C to 40°C
(60.8°F to 104°F),
or under object
temperature range
of 0°C to 100°C
(32°F to 212°F).
The ambient temperature
is above system operation
range, or the object
temperature is above
displayed temperature
range.
The battery cannot provide
enough power for a test.
Replace
the battery
immediately.
The thermometer is broken
or affected by EMF.
Please contact
customer service.
Troubleshooting
Symbol Information
Symbol Referent
Read instructions before use
Manufacturer
Serial number
Authorized representative in the European
Community
Type BF Equipment
Collection for electrical and electronic equipment
Mini Owner’s Manual
Specifications
Memory
20 measurement results
Dimensions
100 x 30.5 x 18 mm
Weight
32 g (include battery)
Power Source
One 3V CR2032 lithium battery
Displayed
temperature range
Forehead: 32°C to 43°C (89.6°F to
109.4°F)
Object: 0°C to 100°C (32°F to 212°F)
Display resolution
0.1°F / 0.1°C
Accuracy
Body mode :
•
±0.2°C (±0.36°F) from 36.0°C (96.8°F)
to 39.0°C (102.2°F)
•
±0.3°C (±0.54°F) from 32.0°C (89.6°F)
to 35.9°C (96.6°F) and from 39.1°C
(102.4°F) to 43.0°C ( 109.4°F)
Object mode :
•
±1°C (±2°F) from 0°C (32°F) to 100°C
(212°F)
Temperature unit
°C or °F
Operating
temperature range
16°C to 40°C (60.8°F to 104°F)
Operating humidity
95% RH or less
Storage
temperature range
-20°C to 50°C (-4°F to 122°F)
Storage humidity
95% RH or less
Mini Owner’s Manual
Device Standard:
Device Corresponds to the requirements of the standard for infrared
thermometers :
• ASTM E1965-98
• EN 60601-1-4 :1996
• EN 12470-5
• EN 60601-1-6:2010
• EN 60601-1-2:2007/AC:2010 • EN 60601-1
Electromagnetic Compatibility :
Device fulfills the stipulations of the standard EN 60601-1-2.
The stipulations of EU-Directive 93/42/EEC for Medical Devices Class Ila have been
fulfilled.
WARRANTY TERMS AND CONDITIONS
With respect to disposable products, Biocare warrants to the original purchaser
that, at time of delivery, each standard product manufactured by Biocare shall be
free of defects in material and workmanship and, when used for the purposes
and indications described on the labeling, is fit for the purposes and indications
described on the labeling. All warranties for a product shall expire as of the
product expiration date, or if none, after 14 months from the date of shipment
from Biocare . Biocare warranty hereunder shall not apply if: (i) a product is not
used in accordance with its instructions or if it is used for a purpose not indicated
on the labeling; (ii) any repairs, alterations or other work has been performed by
Buyer or others on such item, other than work performed with Biocare ’s
authorisation and according to its approved procedures; or (iii) the alleged defect
is a result of abuse, misuse, improper maintenance, accident or the negligence of
any party other than Biocare . The warranty set forth herein is conditioned upon
proper storage, installation, use and maintenance in accordance with applicable
written recommendations of Biocare .
0197
Shanghai International Trading Corp.
GmbH(Hamburg)
Eiffestrasse 80, 20537 Hamburg, Germany
TEL:+49‐40‐2513175
FAX:+49‐40‐255726
Biocare Asia Co., LTD
No.260,Mayun Road, New District Suzhou,
Jiangsu,215129,P.R.C
TEL:+86‐512‐68091772
FAX:+86‐512‐68097761
www.bca-medical.com