6

 

 

Equipment Operation  

Controls and Indicators 

 

INTENDED USERS 

 
The instrument is intended to be used by scientists and technicians who possess basic laboratory 
and technical skills and have the knowledge and understanding of the hazards involved, with the 
unit and the samples used, to operate it in a safe manner. 
 

Instrument Preparation  

 

 

Switch on the unit and allow it to finish its start up calibration 

 

Best performance is obtained if the instrument is allowed to warm up and stabilise for at 
least 30 minutes 

 

If applicable connect the unit to a PC using a USB cable and refer to the online help and user 
manual 

 

Select the appropriate application or method 

 

Where relevant, set up the application parameters for the sample 

 

For Deuterium/ Tungsten units a precision mode is available. To use this mode enter the 
desired application and switch to precision mode 

 

Select cuvette cells to use. It is important to use cells of the correct type. Most samples are 
measured using a standard 10mm path length cell. Special cells and accessories are available 
for larger or smaller path lengths and sample volumes. It is important to use cells of the 

Summary of Contents for Libra S50

Page 1: ...Biochrom Libra S50 S60 S70 S80 UV Visible Spectrophotometers Health Safety Manual...

Page 2: ...ardous Substances in Electrical and Electronic Equipment ROHS 2006 42 EC Machinery directive 1999 5 EC Radio and Telecommunications Terminal Equipment Directive instruments fitted with Bluetooth acces...

Page 3: ...Installation 4 Instrument Connections 5 Equipment Operation 6 Controls and Indicators 6 Intended Users 7 Instrument Preparation 7 Post Run Procedures 7 Performance Validation 7 User Maintenance 7 Tro...

Page 4: ...ard Electrical equipment for measurement control and laboratory use Classified as basic immunity criterion A and Class B for conducted and radiated emissions EN 12100 1 2 2003 A1 2009 Safety of machin...

Page 5: ...sure that the user of the equipment is provided with a safe working environment Any chemicals used in Analyses should be used stored and disposed of in accordance with manufacturer s guidelines and lo...

Page 6: ...hen it should be allowed to come to thermal equilibrium for 2 to 3 hours before operation so that start up calibration is not compromised by condensation The equipment must be connected to the local s...

Page 7: ...5 Instrument Connections...

Page 8: ...m up and stabilise for at least 30 minutes If applicable connect the unit to a PC using a USB cable and refer to the online help and user manual Select the appropriate application or method Where rele...

Page 9: ...nd solvents may be classified as hazardous or bio hazardous waste The disposal of such substances must be carried out in accordance with local regulatory practice Performance Validation Good laborator...

Page 10: ...the online help or user manual TROUBLESHOOTING PROBLEM POSSIBLE CAUSE Start up calibration fails Check sample and reference beam is clear before switch on Possible optical failure contact service supp...

Page 11: ...eb site or your local distributor Return the unit together with a completed declaration of contamination status form Available from the web site or your local distributor Please note that instrumentat...

Page 12: ...10 41 56 1872 issue 01...

Reviews: