Graph
This facility enables scaling of the results, and defines how they are to be presented
on the LCD and print out.
•
Enter the maximum absorbance to be shown on the display during the assay
•
Enter the minimum absorbance to be shown on the display during the assay
•
Select if automatic scaling of the results to fit the display is required post run
•
Select if data points should be indicated on each assay
•
Select if the graph should be printed with the results
Post Run
This facility enables an assay to be selected (in serial or parallel mode) and the slope
to be optimised by redefining the start and end points for the slope.
•
Enter the sample number for optimisation
•
Enter the start point for the slope
•
Enter the end point for the slope
•
Select if these points should be defined automatically
•
If no, the slope is calculated on the basis of the start / end points entered
above
•
Results shown are Initial Abs, Final Abs, Slope, Slope * Factor (result),
Linearity