Issue 02 - 07/2001
English
49
GLP print out at instrument calibration (GLP enabled)
Ultrospec 3300
pro
UV/Vis Spectrophotometer
Lab name ....................
Instrument ..............
Serial no : 81012
Software : 4197 V1.4, Slave 4194 V1.0
Last serviced : 14/08/00 13:36
Instrument state at calibration
GLP Calibrated 09/10/00 at 12:56
Calibration Full UV/Visible
Bandwidth (1.3- 1.8nm): 1.7nm
PASS
Wavelength (656.1nm) : 656.0nm
PASS
Absorbance at
220nm (1.763-1.781A) : 1.772A
PASS
340nm (1.633-1.665A) : 1.649A
PASS
500nm (1.477-1.491A) : 1.484A
PASS
Stray light at
220nm ( <0.025%T) : 0.013%T
PASS
UV lamp : 85 % of original energy
Vis lamp: 95 % of original energy
Current instrument state
Accessory: Eight Position Cell Changer
UV lamp :
installed 14/08/00 13:35, use 5 hours
baseline in use: 14/08/00 13:38
baseline stored: 14/08/00 13:38
Vis lamp :
installed 14/08/00 13:35, use 5 hours
baseline in use: 14/08/00 13:38