Lightwave UV/Vis Diode Array Spectrophotometer
10
Failure of one lamp will not affect the other lamp - so if the Deuterium lamp
fails you can still make good visible measurements (380 to 825 nm).
See Section 12 on Servicing & Maintenance
Failure in Wavelength Calibration
The wavelength calibration is performed by observing the key spectral data
from the Deuterium lamp and ensuring that the peak remains in the correct
position. A failure in wavelength calibration infers some movement of the
fixed optical system or exposure to extreme temperatures.
Try leaving the unit on for a few minutes and then re-start.
If the failure re-occurs please consult your local dealer/service agent or our
service department.
Note: The unit will continue to operate despite being out of wavelength
calibration.
Failure of Diode Array
On start up the unit checks that all pixels (512) are operating. A failure of any
pixel will give the above error.
The unit can still be used as operation at other wavelengths will not be
affected. The faulty pixel will show itself as a line on the wavelength scan.
If the failure occurs please consult your local dealer/service agent or our
service department.