Lightwave UV/Vis Diode Array Spectrophotometer
14
10.3.1 Single measurement
This mode should be used when readout in Absorbance or %Transmission is
required for a single wavelength.
You can toggle between Abs and %T by pressing function key 1.
For very low level Absorbances, variation between cuvettes can be
significant. It is worthwhile to use matched cuvettes, or read both cuvettes
with the reference solution before commencing a series of tests.
10.3.2 Multi Wavelength
The wavelength settings can be changed in a similar manner to the single
measurement mode and are memorised until re-set.
Pressing function key 1 highlights the first wavelength setting, others can be
accessed by pressing the
↑↓
arrows. Pressing function key 1 again allows
the value to be changed (using the
↑↓
arrows). Pressing once more confirms
the choice and moves to the next wavelength.
Once the user is satisfied with the settings press function key 4 “All OK”.
Reference and Test are then active so you can make a measurement.
To avoid accidental changes to the wavelength
setting the arrow keys only become functional
when “Set
λ
“ is pressed. Use the
↑↓
keys to
change wavelength then press “Accept
λ
“ to
accept.