Lightwave UV/Vis Diode Array Spectrophotometer
16
10.4.2 Ratio Measurement
Ratio measurement is used most commonly to measure the purity of DNA
preparations. A value of between 1.7 and 2.0 generally indicates purity.
The wavelength settings can be changed in a similar manner to the single
measurement mode and are memorised until re-set.
Once set, pressing REF or TEST will display the values in Absorbance on the
screen and calculate the Ratio.
10.5 Scanning
The 512-pixel diode array captures readings from the complete wavelength
range each time a measurement is made.
The scan displaying this information can be viewed on the graphics display.
Users may find it useful to look at the spectrum
around 260/280nm to give further indications of
purity. Just press SCAN to have a look.