Lightwave UV/Vis Diode Array Spectrophotometer
26
The communication settings are the same for PC and Printer, but have a
different format. They are accessed from the “Set up instrument” then
“Communications” buttons from the entry menu.
Clicking on the “Device” function key will toggle between Printer and PC.
RS232 settings are as follows. The Baud rate is user adjustable by clicking
on function key 2 “Set Baud” to toggle between the two available values :
Setting Printer/PC
Baud Rate
9600 or 38400
Parity None
Stop Bits
1
Data Bits
8
If connecting to a PC at 38400 Baud the length of the cable should be no
longer than 3m.
11.1.1 Printing
The unit is designed to print to a 40 column serial printer. It is only possible
to print to the WPA S2000P printer using 9600 Baud.
Printer Specifications
Type Serial
RS232
Protocol: Text
EPSON ESC/POS format
Protocol:Graphics
Sends graphics command 1B(hex),01(hex); then 40
bytes of data, the lower 6 bits of each containing the
graphics dots comprising that line. Repeated for
each graphics line of dots.
Width 40
column
Handshaking RTS/CTS
or
Wire in DTR line to the CTS feed from Lightwave
The WPA S2000P printer from WPA meets all the above specifications.