background image

Handheld Pulse Oximeter

17  Environment Setting of Equipment 

Environment Setting of Equipment  18  

PalmCare plus

5.2 Alarm Setting

The below screen will be shown if you press ENTER (          ) button, after making reverse the 

item “1. Alarm Setting” on the environment setting screen of Palmcare plus.

Step1)   The change of the setting value can be controlled by the UP/DOWN  (      /      ), the item 

moving can be controlled ENTER (            )  button.

Step2)   The SETUP screen can be returned if you press the ECS (        ) button  after completed 

inputting Alarm setting.

If you press continuously UP/DOWN (       /       ) button, the value is increased or decreased 

continuously after about one second and the value is increased or decreased by one for each 

on time pressing the UP/DOWN button. The changed alarm range will be displayed on The 

LCD as soon as you set up new alarm range. Please refer the below chart for the allowable 

adjustment range of this.

The LOW+1 in the chart means low limit value +1, the HI-1 means high limit value -1. 

       For example, if low limit value for HR is 55, the HI value range for HR can be adjusted as 

some numbers between 56 to 299. On the contrary, if high limit value for HR is 150, the 

LOW value range for HR can be adjusted as some numbers between1 to 149.

        The value range for SpO

2

 also can be applied of the same rule of setting up HR value. The 

only difference is the high limit value can be “OFF”

        This “OFF” means the measured SpO

2

 value beyond the high limit range will not be checked 

for alarm.

  Push and hold the power button for 1~2 seconds when power off.

  Power off automatically when no inputting data for 1 minute when battery mode.

Step3)   If you want to change the setting value, please press the MENU  (         ) button. Also if you 

want to examine the Trend Data, please press the TREDN (           ) button.

Setting

Default

Value

Range

HI

160
100

HR

SpO

2

Suspend Time

LOW

50
80

1 Min

1 Min / 2 Min / 3 Min, OFF

HI

LOW+1 ~ 299

LOW+1 ~ 100, OFF

LOW

1 ~  HI-1
1 ~  HI-1

5. Environment Setting of Equipment

5.1 Menu Setting

You can set the all menu pushing MENU button (           )

Step1)   If you press MENU button (        )  on the initial screen of Palmcare plus, the environment 

setting mode will be shown as below.

Step2)   Press the ENTER (         ) button and enter setup screen you want, after making reverse turn 

item you want by using UP/DOWN  (         /         ) button. 

Step3)   In case you want to return to the initial screen, press ECS (           ) Button.

Press MENU (            ) Button.

(Alarm setting screen)

NOTE

Summary of Contents for PalmCare plus

Page 1: ...ang gu Seongnam si Gyeonggido 13486 Korea Dir 82 31 696 4899 Fax 82 31 696 4877 Customer Service Dept Dir 82 31 696 4899 Fax 82 31 696 4877 European Representative Medical Econet Gmbh Quality Assuranc...

Page 2: ...OPERATION MANUAL PalmCare plus Handheld Pulse Oximeter Ver 1 3 NOV 10 2017...

Page 3: ...4 Bottom Side 2 5 Back Side 3 Description of Product Symbols 3 1 Screen configuration 4 Palmcare plus Basic Operation 4 1 Operation 3 3 3 4 5 8 9 9 10 10 10 10 11 12 12 12 12 12 12 13 13 16 16 5 Envi...

Page 4: ...special circumstances during the use of the product The warranty is limited to the repair or exchange of the product by the authority of the company or the agency for problems in materials or the manu...

Page 5: ...determining necessity of treatment for the patient WARNING Locations where moisture level increase considerably or where air is not ventilated properly Locations where sudden impact or vibration could...

Page 6: ...battery from the product To ensure the lifecycle and safety of battery please discharge the battery and then charge perfectly again every six months Also time worn battery exceed 24 months should be...

Page 7: ...ttings Pay attention when setting Palmcare plus for the following points Use Palmcare plus under conditions of 10 40 of circumstantial temperature and 30 85 humidity Check the connection status of the...

Page 8: ...ction parts power line or any optional device are connected to the device properly Noise may be made if the device isn t completely grounded Precautions for using Do not disassemble the device Do not...

Page 9: ...for how to set up each initial display mode In order to avoid electrical shock do not open the cover Disassembling of the equipment must be done only an by service personnel who is authorized by BION...

Page 10: ...ds Function of the buttons Each buttons on the PalmCare plus services more than two use according to the screen The use meaning for each screen is as below chart Button Display mode Meaning and Operat...

Page 11: ...56 to 299 On the contrary if high limit value for HR is 150 the LOW value range for HR can be adjusted as some numbers between1 to 149 The value range for SpO2 also can be applied of the same rule of...

Page 12: ...arm ringing when a probe has slipped off or when a probe connector is pulled off from the Palmcare plus At this time this is an item to decide wether this information alarm will activate setting on or...

Page 13: ...end Time Setting 5 3 Initial Display Setting Mode Display Mode Step1 You can enter to the initial display setting mode if you press ENTER button after making a reverse the item 2 DISPLAY MODE on the e...

Page 14: ...Sound Off Volume Off ii 2 Sound Low Low volume iii 3 Sound High High volume Step3 You can return automatically to the SETUP screen if you choose the mode by ENTER button And if you want to cancel the...

Page 15: ...e ENTER button after making a reverse the item 7 Device Info Step2 You can use UP DOWN button for the item moving and ENTER button for your choice Brightness LCD Brightness auto control setting i ON I...

Page 16: ...00 00 for initial trend saving time Saved SpO2 value Save HR Value Event data which is saved by the user Data mark which is saved by Medium PriorityAlarm Data mark which is saved by High PriorityAlarm...

Page 17: ...atal patients make sure the disposable probe should be used outside the incubator The humid atmosphere inside the incubator can cause inaccurate measurements Alarms for Each Mode Standard Mode The hig...

Page 18: ...ower has to be supplied at once For the effective use of battery Palmcare plus has the power saving mode it turn off the LCD back light after 3 seconds and also turn off the all functions of Palmcare...

Page 19: ...act our Customer Service Dept orA S receptions 10 Before Requesting After Service Problem Solution 1 In case of the battery is discharged Connect DC adapter 2 Please check the connection of the switch...

Page 20: ...ation and Pulse rate Memories 72 hours maximum Physical Dimension 165mm H X 70mm W X 30mm D 1CELL 200 g 2CELL 250 g StandardAccessories Finger Probe 5V 2 0ADCAdapter OptionalAccessories USB mini B Typ...

Page 21: ...omer s Damage Compensation Regulation noticed by Economic Planning Board This product is medical equipment Product Warranty Product Name Model Code Manufacture Approval No Manufacture Approval Date Ma...

Reviews: